Ultra‐high temporal resolution 4D angiography using arterial spin labeling with subspace reconstruction
Open Access
- 29 December 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 93 (5) , 1924-1941
Abstract
Purpose: To achieve ultra‐high temporal resolution non‐contrast 4D angiography with improved spatiotemporal fidelity.Methods: Continuous data acquisition using 3D golden‐angle sampling following an arterial spin labeling preparation allows for flexibly reconstructing 4D dynamic angiograms at arbitrary temporal resolutions. However, k‐space data is often temporally “binned” before image reconstruction, negatively affecting spatiotemporal fidelity and limiting temporal resolution. In this work, a subspace was extracted by linearly compressing a dictionary constructed from simulated curves of an angiographic kinetic model. The subspace was used to represent and reconstruct the voxelwise signal timecourse at the same temporal resolution as the data acquisition without temporal binning. Physiological parameters were estimated from the resulting images using a Bayesian fitting approach. A group of eight healthy subjects were scanned and the in vivo results reconstructed by different methods were compared. Because of the difficulty of obtaining ground truth 4D angiograms with ultra‐high temporal resolution, the in vivo results were cross‐validated with numerical simulations.Results: The proposed method enables 4D time‐resolved angiography with much higher temporal resolution (14.7 ms) than previously reported (˜50 ms) while maintaining high spatial resolution (1.1 mm isotropic). Blood flow dynamics were depicted in greater detail, thin vessel visibility was improved, and the estimated physiological parameters also exhibited more realistic spatial patterns with the proposed method.Conclusion: Incorporating a subspace compressed kinetic model into the reconstruction of 4D ASL angiograms notably improved the temporal resolution and spatiotemporal fidelity, which was subsequently beneficial for accurate physiological modeling.Keywords
All Related Versions
Funding Information
- NIHR Oxford Biomedical Research Centre (NIHR203316)
- Royal Academy of Engineering (RF∖201819∖18∖92)
- Canada Research Chairs
- Wellcome Trust (203139/Z/16/Z, 220204/Z/20/Z)
This publication has 39 references indexed in Scilit:
- Noncontrast‐enhanced magnetic resonance angiography and venography imaging with enhanced angiographyJournal of Magnetic Resonance Imaging, 2013
- Temporal stability of adaptive 3D radial MRI using multidimensional golden meansMagnetic Resonance in Medicine, 2009
- Continuous flow‐driven inversion for arterial spin labeling using pulsed radio frequency and gradient fieldsMagnetic Resonance in Medicine, 2008
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraintMagnetic Resonance in Medicine, 2007
- Array compression for MRI with large coil arraysMagnetic Resonance in Medicine, 2007
- Advances in sensitivity encoding with arbitrary k‐space trajectoriesMagnetic Resonance in Medicine, 2001
- k-Space weighted image contrast (KWIC) for contrast manipulation in projection reconstruction MRIMagnetic Resonance in Medicine, 2000
- Adaptive reconstruction of phased array MR imageryMagnetic Resonance in Medicine, 2000
- MR angiography by selective inversion recoveryMagnetic Resonance in Medicine, 1987