Accurate Measurement of Magnetic Resonance Imaging Gradient Characteristics
Open Access
Abstract
Recently, gradient performance and fidelity has become of increasing interest, as the fidelity of the magnetic resonance (MR) image is somewhat dependent on the fidelity of the gradient system. In particular, for high fidelity non-Cartesian imaging, due to non-fidelity of the gradient system, it becomes necessary to know the actual k-space trajectory as opposed to the requested trajectory. In this work we show that, by considering the gradient system as a linear time-invariant system, the gradient impulse response function (GIRF) can be reliably measured to a relatively high degree of accuracy with a simple setup, using a small phantom and a series of simple experiments. It is shown experimentally that the resulting GIRF is able to predict actual gradient performance with a high degree of accuracy. The method captures not only the frequency response but also gradient timing errors and artifacts due to mechanical vibrations of the gradient system. Some discussion is provided comparing the method presented here with other analogous methods, along with limitations of these methods.Keywords
This publication has 22 references indexed in Scilit:
- Gradient system characterization by impulse response measurements with a dynamic field cameraMagnetic Resonance in Medicine, 2012
- Quantification and compensation of eddy-current-induced magnetic-field gradientsJournal of Magnetic Resonance, 2011
- Direct measurement of magnetic field gradient waveformsConcepts in Magnetic Resonance Part A, 2010
- Characterizing and correcting gradient errors in non‐cartesian imaging: Are gradient errors linear time‐invariant (LTI)?Magnetic Resonance in Medicine, 2009
- Estimation of k‐space trajectories in spiral MRIMagnetic Resonance in Medicine, 2009
- Improved k‐space trajectory measurement with signal shiftingMagnetic Resonance in Medicine, 2007
- A Method to measure arbitrary k‐space trajectories for rapid MR imagingMagnetic Resonance in Medicine, 1997
- Quantitative characterization of the eddy current fields in a 40‐cm bore superconducting magnetMagnetic Resonance in Medicine, 1994
- A simple method of measuring gradient induced eddy currents to set compensation networksMagnetic Resonance in Medicine, 1993
- Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systemsJournal of Magnetic Resonance (1969), 1990