Variable‐flip‐angle 3D spiral‐in‐out turbo spin‐echo imaging using concomitant gradient compensation and echo reordering at 0.55 T
Open Access
- 20 November 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 93 (4) , 1741-1750
Abstract
Purpose: To develop single‐slab 3D spiral turbo spin echo (spiral SPACE) for 1‐mm 3 isotropic whole‐brain T 2 ‐weighted imaging on a high‐performance 0.55T scanner, with high scan efficiency from interleaved spiral‐in‐out trajectories, variable‐flip‐angle refocusing radiofrequency (RF) pulses, echo reordering, and concomitant‐field compensation. Methods: A stack‐of‐spirals (in‐out waveforms) turbo‐spin‐echo acquisition was implemented with T 2 ‐weighed contrast. Gradient infidelity was corrected using the gradient impulse response function (GIRF), and concomitant‐field compensation was used to correct for phase errors among echoes and during the readout windows. To maintain a long echo train (˜600 ms) within each shot, variable‐flip‐angle refocusing RF pulses were generated using extended‐phase‐graph analysis. An echo‐reordering scheme provided a smooth signal variation along the echo direction in k‐space. Images from spiral SPACE with and without concomitant‐field compensation were compared with those from Cartesian SPACE in phantoms and 6 healthy volunteers. Results: Phantom results demonstrated the improved performance of concomitant‐field correction via sequence‐based modifications and of GIRF–based trajectory estimation. Volunteer data showed that with concomitant‐field correction and echo reordering, system imperfection associated image artifacts and blurring were substantially mitigated in spiral SPACE. Compared with Cartesian SPACE, spiral SPACE had an overall 15%–25% signal‐to‐noise ratio (SNR) improvement in both white matter and gray matter. Conclusion: A 3D spiral‐in‐out SPACE acquisition with variable‐flip‐angles, concomitant‐field compensation, and echo‐reordering was demonstrated at 0.55 T, showing promising gains in SNR, compared with Cartesian SPACE.Keywords
Funding Information
- Division of Intramural Research (Z1A‐HL006257)
- Siemens Medical Solutions USA (UVA‐2018‐MR‐27‐01‐Mugler_C00227955)
- Siemens Medical Solutions USA (UVA‐2021‐MR‐29‐01‐Mugler_C00239406)
- National Institute of Biomedical Imaging and Bioengineering (R01 EB028773)
This publication has 19 references indexed in Scilit:
- Real‐time distortion correction of spiral and echo planar images using the gradient system impulse response functionMagnetic Resonance in Medicine, 2015
- Sliding‐slab three‐dimensional TSE imaging with a spiral‐In/Out readoutMagnetic Resonance in Medicine, 2015
- Optimized three‐dimensional fast‐spin‐echo MRIJournal of Magnetic Resonance Imaging, 2014
- ESPIRiT—an eigenvalue approach to autocalibrating parallel MRI: Where SENSE meets GRAPPAMagnetic Resonance in Medicine, 2013
- Gradient system characterization by impulse response measurements with a dynamic field cameraMagnetic Resonance in Medicine, 2012
- Comprehensive quantification of signal‐to‐noise ratio and g‐factor for image‐based and k‐space‐based parallel imaging reconstructionsMagnetic Resonance in Medicine, 2008
- Fast spin echo sequences with very long echo trains: Design of variable refocusing flip angle schedules and generation of clinical T2 contrastMagnetic Resonance in Medicine, 2006
- Calculation of flip angles for echo trains with predefined amplitudes with the extended phase graph (EPG)‐algorithm: Principles and applications to hyperecho and TRAPS sequencesMagnetic Resonance in Medicine, 2003
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Artifacts induced by concomitant magnetic field in fast spin‐echo imagingMagnetic Resonance in Medicine, 1998