Abstract
Purpose: To develop single‐slab 3D spiral turbo spin echo (spiral SPACE) for 1‐mm 3 isotropic whole‐brain T 2 ‐weighted imaging on a high‐performance 0.55T scanner, with high scan efficiency from interleaved spiral‐in‐out trajectories, variable‐flip‐angle refocusing radiofrequency (RF) pulses, echo reordering, and concomitant‐field compensation. Methods: A stack‐of‐spirals (in‐out waveforms) turbo‐spin‐echo acquisition was implemented with T 2 ‐weighed contrast. Gradient infidelity was corrected using the gradient impulse response function (GIRF), and concomitant‐field compensation was used to correct for phase errors among echoes and during the readout windows. To maintain a long echo train (˜600 ms) within each shot, variable‐flip‐angle refocusing RF pulses were generated using extended‐phase‐graph analysis. An echo‐reordering scheme provided a smooth signal variation along the echo direction in k‐space. Images from spiral SPACE with and without concomitant‐field compensation were compared with those from Cartesian SPACE in phantoms and 6 healthy volunteers. Results: Phantom results demonstrated the improved performance of concomitant‐field correction via sequence‐based modifications and of GIRF–based trajectory estimation. Volunteer data showed that with concomitant‐field correction and echo reordering, system imperfection associated image artifacts and blurring were substantially mitigated in spiral SPACE. Compared with Cartesian SPACE, spiral SPACE had an overall 15%–25% signal‐to‐noise ratio (SNR) improvement in both white matter and gray matter. Conclusion: A 3D spiral‐in‐out SPACE acquisition with variable‐flip‐angles, concomitant‐field compensation, and echo‐reordering was demonstrated at 0.55 T, showing promising gains in SNR, compared with Cartesian SPACE.
Funding Information
  • Division of Intramural Research (Z1A‐HL006257)
  • Siemens Medical Solutions USA (UVA‐2018‐MR‐27‐01‐Mugler_C00227955)
  • Siemens Medical Solutions USA (UVA‐2021‐MR‐29‐01‐Mugler_C00239406)
  • National Institute of Biomedical Imaging and Bioengineering (R01 EB028773)

This publication has 19 references indexed in Scilit: