The Impact of Combining a Low-Tube Voltage Acquisition with Iterative Reconstruction on Total Iodine Dose in Coronary CT Angiography
Open Access
- 23 May 2017
- journal article
- research article
- Published by Wiley in BioMed Research International
- Vol. 2017, 1-10
Abstract
Objectives. To assess the impact of combining low-tube voltage acquisition with iterative reconstruction (IR) techniques on the iodine dose in coronary CTA. Methods. Three minipigs underwent CCTA to compare a standard of care protocol with two alternative study protocols combining low-tube voltage and low iodine dose with IR. Image quality was evaluated objectively by the CT value, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the main coronary arteries and aorta and subjectively by expert reading. Statistics were performed by Mann–Whitney test and Chi-square analysis. Results. Despite reduced iodine dose, both study protocols maintained CT values, SNR, and CNR compared to the standard of care protocol. Expert readings confirmed these findings; all scans were perceived to be of at least diagnostically acceptable quality on all evaluated parameters allowing image interpretation. No statistical differences were observed (all values > 0.11), except for streak artifacts () which were considered to be more severe, although acceptable, with the 80 kVp protocol. Conclusions. Reduced tube voltage in combination with IR allows a total iodine dose reduction between 37 and 50%, by using contrast media with low iodine concentrations of 200 and 160 mg I/mL, while maintaining image quality.
Keywords
Funding Information
- GE Healthcare Medical Diagnostics
This publication has 32 references indexed in Scilit:
- CT Angiography of the Head-and-Neck Vessels Acquired with Low Tube Voltage, Low Iodine, and Iterative Image Reconstruction: Clinical Evaluation of Radiation Dose and Image QualityPLOS ONE, 2013
- Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT AngiographyInvestigative Radiology, 2013
- Contrast-induced kidney injury: mechanisms, risk factors, and preventionEuropean Heart Journal, 2012
- Coronary Enhancement for Prospective ECG-Gated Single R-R Axial 320-MDCT Angiography: Comparison of 60- and 80-mL Iopamidol 370 InjectionAmerican Journal of Roentgenology, 2011
- Radiation dose reduction with increasing utilization of prospective gating in 64-multidetector cardiac computed tomography angiographyJournal of Cardiovascular Computed Tomography, 2011
- Toxic effects of a high dose of non-ionic iodinated contrast media on renal glomerular and aortic endothelial cells in aged rats in vivoToxicology Letters, 2011
- Adaptive Statistical Iterative Reconstruction: Assessment of Image Noise and Image Quality in Coronary CT AngiographyAmerican Journal of Roentgenology, 2010
- Low-Tube-Voltage, High-Tube-Current Multidetector Abdominal CT: Improved Image Quality and Decreased Radiation Dose with Adaptive Statistical Iterative Reconstruction Algorithm—Initial Clinical ExperienceRadiology, 2010
- Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility StudyAmerican Journal of Roentgenology, 2009
- The prognostic implications of further renal function deterioration within 48 h of interventional coronary procedures in patients with pre-existent chronic renal insufficiencyJACC, 2000