Programmable Readout With Integrated Bandgap Reference Potentiostat for Glucose Sensing
- 23 February 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 73 (00189456) , 1-11
Abstract
A novel electronics interface which includes programmable readout amplifier and bandgap reference voltage potentiostat circuit for glucose electrochemical sensors is designed and implemented in 0.18 μm, CMOS process technology and 1.5 V supply voltage. One of the main advantage of proposed architecture is that it has digitally controlled readout circuit and a bandgap reference voltage potentiostat integrated into a single circuit. The proposed readout circuit i.e the programmable transimpedance amplifier (PTIA) offers wide dynamic range and low noise. The PTIA achieves an overall transimpedance gain of 17.3-50.5 kΩ . The linear output voltage range of the PTIA response is 0.55-1.44 V for an input current range of 4.2-180 μA. The output rms noise value is obtained to be 5.101 μV rms and the complete architecture dissipates a total power of 2.33 mW. The % THD varying from 0.5 to 4 for the above specified current range. A 0.6 V reference voltage has been used for all measurements produced by bandgap reference voltage potentiostat. A glassy carbon electrode (GCE) filled-in coated with CuO/Cu 0.76 CO 2.25 O 4 (copper cobaltite) is used as the working electrode. The glucose concentration ranges from 1 mM to 10 mM has been measured by electrochemical glucose sensing setup and the prototyped circuit tested for the practicability of proposed architecture. The total area of the proposed glucose sensing architecture is 0.0684 mm 2 .Keywords
Funding Information
- Govt. of India and World Bank (TEQIPIII)
- Technology Development Programme of IITI DRISHTI CPS Foundation
- National Mission on Interdisciplinary Cyber Physical System
- Department of Science and Technology, Government of India
This publication has 35 references indexed in Scilit:
- Novel tunable current feedback instrumentation amplifier based on BBFC OP-AMP for biomedical applications with low power and high CMRRIntegration, 2023
- A low-noise and mismatch-tolerant current-mirror-based potentiostat circuit for glucose monitoringMicroelectronics Journal, 2023
- Machine learning models for non-invasive glucose measurement: towards diabetes management in smart healthcareHealth and Technology, 2022
- Design of potentiostat and current mode read-out amplifier for glucose sensing2021
- A low-power low-noise multi-stage transimpedance amplifier for amperometric based blood glucose monitoring systemsAnalog Integrated Circuits and Signal Processing, 2020
- Amperometric Biosensor and Front-End Electronics for Remote Glucose Monitoring by Crosslinked PEDOT-Glucose OxidaseIEEE Sensors Journal, 2018
- Low power, low noise, compact amperometric circuit for three-terminal glucose biosensorAnalog Integrated Circuits and Signal Processing, 2016
- A CMOS Analog Correlator-Based Painless Nonenzymatic Glucose Sensor Readout CircuitIEEE Sensors Journal, 2014
- A Fully Differential PotentiostatIEEE Sensors Journal, 2009
- A comparison between potentiostatic circuits with grounded work or auxiliary electrodeReview of Scientific Instruments, 2002