Abstract
Propane in nitrogen (C3H8/N2) gas mixtures are widely used as calibration standards for hydrocarbon measurements in environmental monitoring, energy production and industrial process control. Five C3H8/N2 gas mixtures, covering the range of 804.8148-12053.8148 µmol/mol, were gravimetrically prepared according to ISO 6142 using a two-step dilution approach with high-purity propane and nitrogen. The prepared mixtures were validated by gas chromatography with flame ionization detection (GC-FID) in accordance with ISO 6143. The GC-FID results showed good agreement with the gravimetric values, confirming the validity of the preparation method. The certified values determined by gravimetry were established for 4 mixtures (B, C, D, E) as 12053.81 ± 97.49, 9013.23 ± 78.11, 5998.011 ± 57.72 and 2987.29 ± 66.91 µmol/mol, respectively. The mole fraction of the mixture F, 804.81 ± 35.97 µmol/mol was considered an information value since it was associated with large uncertainty of 7.016%. The performance of the produced certified reference materials (CRMs) was checked by using them to calibrate the GC-FID, then measure a CRM sample of 8010 µmol/mol as an unknown. The precision (%RSD) and accuracy of the measured results were found to be 0.12% and 99.75%, respectively. These results confirm that the prepared C3H8/N2 mixtures meet the criteria for certified reference materials and are suitable as calibrants for hydrocarbon emission measurement instruments in industrial gas analysis and environmental monitoring.