Investigating photomultiplier tube nonlinearities in high-speed phosphor thermometry using light emitting diode simulated decay curves
Open Access
- 1 December 2021
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 92 (12) , 123102
Abstract
Photomultiplier tube (PMT) nonlinearities relevant for single shot high-speed lifetime phosphor thermometry were investigated by simulating decay curves with a light emitting diode (LED) at repetition rates between 1 Hz and 10 kHz. The PMT gain, LED decay time, and background radiant flux were also varied to investigate their impact on the measured decay time error. Errors in the measured decay time due to nonlinear PMT performance lead to temperature measurement errors; therefore, having the measured decay time sensitive to only phosphor temperature is highly valuable for more reliable temperature measurements. Photocathode bleaching had a significant impact on the signal level linearity for PMTs with excitation frequency in the kHz regime but had a smaller impact on the decay time error. Space charge effects were most noticeable at high radiant flux levels and high repetition rates. Strong background radiant flux may lead to decay time errors, and a gateable photocathode could be an effective method to reduce decay time errors. The best decay time measurement configuration to maximize precision without sacrificing accuracy is to use PMT gain in the recommended range and the highest radiant flux where the PMT response is still linear. The degree of nonlinearity in the PMT response is partly detector dependent; therefore, the results in this work may differ among detectors; however, the analysis presented in this work provides guidelines for improving the temperature accuracy of kHz lifetime phosphor thermometry measurements.Keywords
Funding Information
- Energimyndigheten (45400-01)
- European Commission (884229)
This publication has 18 references indexed in Scilit:
- Phosphor thermometry: A comparison of the luminescence lifetime and the intensity ratio approachProceedings of the Combustion Institute, 2013
- Comparison of photo detectors and operating conditions for decay time determination in phosphor thermometryReview of Scientific Instruments, 2012
- Investigation and compensation of the nonlinear response in photomultiplier tubes for quantitative single-shot measurementsReview of Scientific Instruments, 2012
- High-speed phosphor thermometryReview of Scientific Instruments, 2011
- Effects of incomplete decay in fluorescence lifetime estimationBiomedical Optics Express, 2011
- Progress towards absolute intensity measurements of emissions from high temperature thermographic phosphorsJournal of Luminescence, 2011
- An algorithm for the characterisation of multi-exponential decay curvesOptics and Lasers in Engineering, 2009
- Phosphor Thermometry in Combustion ApplicationsApplied Spectroscopy, 1999
- Remote thermometry with thermographic phosphors: Instrumentation and applicationsReview of Scientific Instruments, 1997