HIFU Drive System Miniaturization Using Harmonic Reduced Pulsewidth Modulation
Open Access
- 29 October 2018
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 65 (12) , 2407-2417
Abstract
Switched excitation has the potential to improve on the cost, efficiency, and size of the linear amplifier circuitry currently used in high-intensity focused ultrasound (HIFU) systems. Existing switching schemes are impaired by high harmonic distortion or lack array apodisation capability, so require adjustable supplies and/or large power filters to be useful. A multilevel pulsewidth modulation (PWM) topology could address both of these issues but the switching-speed limitations of transistors mean that there are a limited number of pulses available in each waveform cycle. In this study, harmonic reduction PWM (HRPWM) is proposed as an algorithmic solution to the design of switched waveforms. Its appropriateness for HIFU was assessed by design of a high power five-level unfiltered amplifier and subsequent thermal-only lesioning of ex vivo chicken breast. Three switched waveforms of different electrical powers (16, 26, 35 W) were generated using the HRPWM algorithm. Lesion sizes were measured and compared with those made at the same electrical power using a linear amplifier and bi-level excitation. HRPWM produced symmetric, thermal-only lesions that were the same size as their linear amplifier equivalents (p > 0.05). At 16 W, bi-level excitation produced smaller lesions but at higher power levels large transients in the acoustic waveform nucleated undesired cavitation. These results demonstrate that HRPWM can minimize HIFU drive circuity size without the need for filters to remove harmonics or adjustable power supplies to achieve array apodisation.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/K029835/1)
- National Heart, Lung, and Blood Institute (1 U01 HL121838-01)
- Wellcome Trust (105615/Z/14/Z)
- Royal Society (RG170324)
This publication has 67 references indexed in Scilit:
- Thresholds for thermal damage to normal tissues: An updateInternational Journal of Hyperthermia, 2011
- Real-Time Monitoring of High-Intensity Focused Ultrasound Lesion Formation Using Acousto-Optic SensingUltrasound in Medicine & Biology, 2011
- Guidance on Reporting Ultrasound Exposure Conditions for Bio-Effects StudiesUltrasound in Medicine & Biology, 2011
- Driving Circuitry for Focused Ultrasound Noninvasive Surgery and Drug Delivery ApplicationsSensors, 2011
- Staircase-Voltage Metal–Oxide–Semiconductor Field-Effect Transistor Driver Circuit for Therapeutic UltrasoundJapanese Journal of Applied Physics, 2010
- Acoustic standing wave suppression using randomized phase-shift-keying excitationsThe Journal of the Acoustical Society of America, 2009
- Acoustic characterization of high intensity focused ultrasound fields: A combined measurement and modeling approachThe Journal of the Acoustical Society of America, 2008
- An image‐guided high intensity focused ultrasound device for uterine fibroids treatmentMedical Physics, 2002
- Ultrasonic heating of the skullThe Journal of the Acoustical Society of America, 1990
- Consequences of Failure to Meet Assumptions Underlying the Fixed Effects Analyses of Variance and CovarianceReview of Educational Research, 1972