Abstract
Accurate calibration of atomic magnetometers and the detection of weak magnetic signals in applications such as biomagnetism and fundamental physics experiments are hindered by low-frequency environmental disturbances that mask useful signals. To address this, we implement a magnetic field reference source system that integrates an NMOR atomic magnetometer with a predictive-control-based active compensation loop. The system establishes a quiet and stable magnetic environment, supporting both magnetometer calibration and weak-field measurements. By modeling the NMOR response, the relation between field variation and demodulated phase is derived and used as a robust error signal for stabilization. Under a 10000 nT bias field, the system suppresses low-frequency noise by 42.6 dB compared with passive shielding and achieves residual levels of 264 fT/√Hz at 0.1 Hz. Allan deviation analysis further verifies improved long-term stability, enabling accurate biomagnetic and low-field NMR studies as well as performance evaluation of precision magnetometers.