Validating Fricke dosimetry for the measurement of absorbed dose to water for HDR 192 Ir brachytherapy: a comparison between primary standards of the LCR, Brazil, and the NRC, Canada
Open Access
- 1 April 2018
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (8) , 085004
Abstract
Two Fricke-based absorbed dose to water standards for HDR Ir-192 dosimetry, developed independently by the LCR in Brazil and the NRC in Canada have been compared. The agreement in the determination of the dose rate from a HDR Ir-192 source at 1 cm in a water phantom was found to be within the k = 1 combined measurement uncertainties of the two standards: D NRC / D LCR = 1.011, standard uncertainty = 2.2%. The dose-based standards also agreed within the uncertainties with the manufacturer’s stated dose rate value, which is traceable to a national standard of air kerma. A number of possible influence quantities were investigated, including the specific method for producing the ferrous-sulphate Fricke solution, the geometry of the holder, and the Monte Carlo code used to determine correction factors. The comparison highlighted the lack of data on the determination of G (Fe 3+ ) in this energy range and the possibilities for further development of the holders used to contain the Fricke solution. The comparison also confirmed the suitability of Fricke dosimetry for Ir-192 primary standard dose rate determinations at therapy dose levels.Keywords
This publication has 23 references indexed in Scilit:
- Comparison BIPM.RI(I)-K8 of high dose-rate Ir-192 brachytherapy standards for reference air kerma rate of the NMIJ and the BIPMMetrologia, 2016
- The Fricke dosimeter as an absorbed dose to water primary standard for Ir-192 brachytherapyPhysics in Medicine & Biology, 2015
- A Feasibility Study of Fricke Dosimetry as an Absorbed Dose to Water Standard for 192Ir HDR SourcesPLOS ONE, 2014
- Using LiF:Mg,Cu,P TLDs to estimate the absorbed dose to water in liquid water around an192Ir brachytherapy sourceMedical Physics, 2013
- A dosimetric uncertainty analysis for photon‐emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC‐ESTROMedical Physics, 2011
- Air kerma standard for calibration of well‐type chambers in Brazil using HDR sources and its traceabilityMedical Physics, 2009
- A package of Linux scripts for the parallelization of Monte Carlo simulationsComputer Physics Communications, 2006
- Spectra and air-kerma strength for encapsulated 192Ir sourcesMedical Physics, 1999
- Calibration of 192Ir high‐dose‐rate afterloading systemsMedical Physics, 1991
- Calibration of the Ferrous Sulphate Dosimeter by Ionometric and Calorimetric MethodsNature, 1966