Registration accuracy with the low dose kilovoltage cone-beam CT: A phantom study
Open Access
- 1 July 2019
- journal article
- Published by Oxford University Press (OUP) in BJR|Open
- Vol. 1 (1)
Abstract
The aim of this study was to investigate low-dose kilovoltage cone-beam CT (kV-CBCT) for image-guided radiotherapy, with a particular focus on the accuracy of image registration with low-dose protocols. Imaging doses were measured with a NOMEX semiconductor detector positioned at the front of head, thorax, and pelvis human body phantoms while kV-CBCT scans were acquired at different tube currents. Aspects of image quality (spatial resolution, noise, uniformity, contrast, geometric distortion, and Hounsfield unit sensitivity) and image registration accuracy using bone and soft tissue were evaluated. With preset and the lowest tube currents, the imaging doses were 0.16 and 0.08 mGy, 5.29 and 2.80 mGy, and 18.23 and 2.69 mGy for head, thorax, and pelvis, respectively. Noise was the only quality aspect directly dependent on tube current, being increased by 1.5 times with a tube current half that of the preset in head and thorax, and by 2.2 times with a tube current 1/8 of the preset in the pelvis. Accurate auto-bone matching was performed within 1 mm at the lowest tube current. The auto-soft tissue matching could not be performed with the lowest tube current; however, manual-soft tissue matching could still be performed within 2 mm or less. Noise was the only image quality aspect dependent on the imaging dose. Auto-bone and manual-soft tissue matching could still be performed at the lowest imaging dose. When optimizing kV-CBCT imaging dose, the impact on bone and soft tissue image registration accuracy should be evaluated.Keywords
This publication has 17 references indexed in Scilit:
- AAPM TG158: Measurement and calculation of doses outside the treated volume from external‐beam radiation therapyMedical Physics, 2017
- Cone beam CT dose reduction in prostate radiotherapy using Likert scale methodsBritish Journal of Radiology, 2016
- Imaging dose from cone beam computed tomography in radiation therapyPhysica Medica, 2015
- Radiochromic film based dosimetry of image‐guidance procedures on different radiotherapy modalitiesJournal of Applied Clinical Medical Physics, 2014
- Dose Distribution from kV-Cone Beam Computed Tomography in Image-guided RadiotherapyJapanese Journal of Radiological Technology, 2013
- Personalized Assessment of kV Cone Beam Computed Tomography Doses in Image-guided Radiotherapy of Pediatric Cancer PatientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Radiation-Related Risk of Basal Cell Carcinoma: A Report From the Childhood Cancer Survivor StudyJNCI: Journal of the National Cancer Institute, 2012
- Evaluation of Radiation Dose and Image Quality for the Varian Cone Beam Computed Tomography SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Assessment of patient organ doses and effective doses using the VIP-Man adult male phantom for selected cone-beam CT imaging procedures during image guided radiation therapyRadiation Protection Dosimetry, 2008
- The management of imaging dose during image‐guided radiotherapy: Report of the AAPM Task Group 75Medical Physics, 2007