Deriving target exposure indices for common radiography exams based on automatic exposures of flat‐field phantoms
Open Access
- 28 April 2026
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 27 (5) , e70562
Abstract
Background: The IEC exposure index (EI), deviation index (DI), and target exposure index (EI T ), represent critical standardized metrics for the evaluation of exposure and quality in radiographic imaging. Purpose: This work develops and validates a systematic procedure to estimate the EI T for eight of the most common radiography imaging protocols utilizing automatic exposure control (AEC) from measurements acquired under reference conditions. Methods: A model was developed to define the relationship between a systems AEC logic, and an estimation of the EI under flat‐field conditions (EIFFC). Separately, clinical data and acquisition protocol information for resultant EI during patient studies were also collected for the eight protocols studied: Chest posteroanterior (PA), Chest lateral, Abdomen anteroposterior (AP), Pelvis AP, L‐Spine AP, C‐Spine AP, T‐Spine AP, and Ribs AP. Data were collected from 41 x‐ray units spanning seven institutions. For each protocol on each unit the EIFFC was computed based on the acquisition protocol, as well as median EI from clinical exams to produce a scaling factor (SF). Kruskal–Wallis statistical tests were used to compare SF's between vendors and AEC cell configurations. Results: SFs for eight radiographic imaging protocols have been produced per vendor and per AEC cell selection. A workflow has been established for end‐users to follow to apply these SFs to flat‐field measurements taken at their own locations to establish local EI T . Conclusions: The study results show that in seven of the eight imaging protocols, the SFs for most units included in the study report SFs within 1 DI (± 25%) of their respective final vendor reported SF (218/228). The ribs protocol is the exception to this finding (n = 26). SFs have high utility for establishing EI T values on individual x‐ray units and normalizing EI value distributions for quality assurance purposes.Keywords
This publication has 10 references indexed in Scilit:
- ACPSEM position paper: recommendations for a digital general X-ray quality assurance programPhysical and Engineering Sciences in Medicine, 2024
- Quality management, quality assurance, and quality control in medical physicsJournal of Applied Clinical Medical Physics, 2023
- Optimising default radiographic exposure factors using Deviation IndexRadiography, 2020
- Current state of practice regarding digital radiography exposure indicators and deviation indices: Report of AAPM Imaging Physics Committee Task Group 232Medical Physics, 2018
- Automatic exposure control calibration and optimisation for abdomen, pelvis and lumbar spine imaging with an Agfa computed radiography systemPhysics in Medicine & Biology, 2016
- Image Gently Campaign Back to Basics Initiative: Ten Steps to Help Manage Radiation Dose in Pediatric Digital RadiographyAmerican Journal of Roentgenology, 2013
- Quality assurance: using the exposure index and the deviation index to monitor radiation exposure for portable chest radiographs in neonatesPediatric Radiology, 2010
- Calibrating automatic exposure control devices for digital radiographyPhysics in Medicine & Biology, 2006
- Optimising automatic exposure control in computed radiography and the impact on patient doseRadiation Protection Dosimetry, 2005
- The Current Concept of Speed Should Not Be Used to Describe Digital Imaging SystemsRadiology, 2005