A new, ultra-stable lab for nano-metrology of x-ray optics at Diamond Light Source
- 1 October 2025
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 96 (10)
Abstract
A new Optics Metrology Laboratory for assembling and characterizing beamline x-ray optical systems has been built. This replaces the old laboratory, which was demolished to make space for construction of a new flagship beamline for the forthcoming Diamond-II facility upgrade. The new cleanroom laboratory is located between several beamlines and laboratories, which intermittently generate significantly higher levels of acoustic noise and floor vibrations. A threefold design strategy was employed to create an ultra-stable environment for the sensitive, optical metrology instruments. First, the walls, ceiling and doors of the laboratory were constructed to attenuate acoustic noise. Second, the air handling systems were designed to minimize self-production of noise and vibrations. Finally, engineering solutions were developed to further isolate the metrology instruments from environmental fluctuations. Overall, despite higher levels of external disturbances, this strategy enables nano-metrology to be successfully conducted in the new laboratory. The shielded environment around each instrument achieves noise rating NR30, which is 5–25 dB quieter than the old laboratory. Over 60-h, the temperature inside the Diamond-NOM’s enclosure varied by only 0.004 °C rms, and humidity changed by <1% RH. All optical metrology instruments are now performing better than in the old laboratory: the slope error repeatability of Diamond-NOM is improved from 15 to 9 nrad rms; the GTX micro-interferometer has measured super-polished substrates with micro-roughness <40 pm rms; the new gantry for Speckle Angular Measurement is commissioned; and the HDX Fizeau interferometer has measured mirrors with slope errors <50 nrad.Keywords
Funding Information
- Diamond Light Source
This publication has 37 references indexed in Scilit:
- A Fizeau interferometry stitching system to characterize X-ray mirrors with sub-nanometre errorsOptics and Lasers in Engineering, 2022
- Nano-precision metrology of X-ray mirrors with laser speckle angular measurementLight: Science & Applications, 2021
- Absolute calibration of optical flats using the three-flat test by considering the relative humidity changeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Development of a high-performance gantry system for a new generation of optical slope measuring profilersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Characterization of the error budget of Alba-NOMNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- The upgraded LTP-V at SLSNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Current status of the NSLS-II optical metrology laboratoryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Sub-microradian surface slope metrology with the ALS Developmental Long Trace ProfilerNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Characterization and calibration of 2nd generation slope measuring profilerNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Modern Developments in X-Ray and Neutron Optics2008