Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review
- 1 August 2017
- journal article
- review article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 90 (1076) , 20160406
Abstract
This article reviews publications related to the use of CT scans for radiotherapy treatment planning, specifically the impact of scan protocol changes on CT number and treatment planning dosimetry and on CT image quality. A search on PubMed and EMBASE and a subsequent review of references yielded 53 relevant articles. CT scan parameters significantly affect image quality. Some will also affect Hounsfield unit (HU) values, though this is not comprehensively reported on. Changes in tube kilovoltage and, on some scanners, field of view and reconstruction algorithms have been found to produce notable HU changes. The degree of HU change which can be tolerated without changing planning dose by >1% depends on the body region and size, planning algorithms, treatment beam energy and type of plan. A change in soft-tissue HU value has a greater impact than changes in HU for bone and air. The use of anthropomorphic phantoms is recommended when assessing HU changes. There is limited published work on CT scan protocol ...Keywords
This publication has 59 references indexed in Scilit:
- Calibration of CT Hounsfield units for proton therapy treatment planning: use of kilovoltage and megavoltage images and comparison of parameterized methodsPhysics in Medicine & Biology, 2013
- Automated delineation of radiotherapy volumes: are we going in the right direction?British Journal of Radiology, 2013
- Dosimetric comparison of different inhomogeneity correction algorithms for external photon beam dose calculationsJournal of Medical Physics, 2013
- Dosimetric verification of radiotherapy treatment planning systems in Serbia: national auditRadiation Oncology, 2012
- Extracting atomic numbers and electron densities from a dual source dual energy CT scanner: Experiments and a simulation modelRadiotherapy and Oncology, 2011
- The effects of field-of-view and patient size on CT numbers from cone-beam computed tomographyPhysics in Medicine & Biology, 2009
- Variability of Target and Normal Structure Delineation for Breast Cancer Radiotherapy: An RTOG Multi-Institutional and Multiobserver StudyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Optimal acquisition parameter selection for CT simulators in radiation oncologyJournal of Applied Clinical Medical Physics, 2008
- Commissioning a new CT simulator I: CT simulator hardwareJournal of Radiotherapy in Practice, 2007
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996