The effect of device configuration and patient’s body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T
Preprint
Abstract
BACKGROUND: Patients with deep brain stimulation (DBS) implants have limited access to MRI due to safety concerns associated with RF-induced heating. Currently, MRI in these patients is allowed only in 1.5T horizontal scanners and with pulse sequences with reduced power. Nevertheless, off-label use of MRI at 3T is increasingly reported based on limited safety assessments. Here we present results of systematic RF heating measurements for two commercially available DBS systems during MRI at 1.5T and 3T. PURPOSE: To assess the effect of imaging landmark, DBS lead configuration, and patient body composition on RF heating of DBS leads during MRI at 1.5 T and 3T. STUDY TYPE: Phantom study. POPULATION/SUBJECTS/PHANTOM/SPECIMEN/ANIMAL MODEL: Gel phantoms and cadaver brain. FIELD STRENGTH/SEQUENCE: 1.5T and 3T, T1-weighted turbo spin echo. ASSESSMENT: RF heating was measured at tips of DBS leads implanted in brain-mimicking gel. STATISTICAL TESTS: None. RESULTS: We observed substantial fluctuation in RF heating mainly affected by phantom composition and DBS lead configuration, ranging from 0.14°C to 23.73°C at 1.5 T, and from 0.10°C to 7.39°C at 3T. The presence of subcutaneous fat substantially altered RF heating at electrode tips (−3.06°C < Δ T < 19.05°C). Introducing concentric loops in the extracranial portion of the lead at the surgical burr hole reduced RF heating by up to 89% at 1.5T and up to 98% at 3T compared to worst case heating scenarios. DATA CONCLUSION: Device configuration and patient body composition significantly altered the RF heating of DBS leads during MRI at 1.5T and 3T. Interestingly, certain lead trajectories consistently reduced RF heating and image artifact over different imaging landmarks, RF frequencies, and phantom compositions. Such trajectories could be implemented in patients with minimal disruption to the surgical workflow.Keywords
All Related Versions
- Published version: Journal of Magnetic Resonance Imaging, 53 (2), 599.
This publication has 30 references indexed in Scilit:
- Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loopsMagnetic Resonance in Medicine, 2016
- Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implantsMagnetic Resonance in Medicine, 2016
- Safely assessing radiofrequency heating potential of conductive devices using image‐based current measurementsMagnetic Resonance in Medicine, 2014
- 3D Slicer as an image computing platform for the Quantitative Imaging NetworkMagnetic Resonance Imaging, 2012
- Deep brain stimulation of the subthalamic nucleus for the treatment of Parkinson's diseaseThe Lancet Neurology, 2009
- Pallidal Deep-Brain Stimulation in Primary Generalized or Segmental DystoniaNew England Journal of Medicine, 2006
- Bilateral deep brain stimulation in Parkinson's disease: a multicentre study with 4 years follow-upBrain, 2005
- Deep Brain Stimulation for Treatment-Resistant DepressionNeuron, 2005
- Reduction of resonant RF heating in intravascular catheters using coaxial chokesMagnetic Resonance in Medicine, 2000
- Nonsusceptibility artifacts due to metallic objects in MR imagingJournal of Magnetic Resonance Imaging, 1995