Abstract
Purpose: To quantify the angular dependence of monofocal intraocular lens (IOL) power. Setting: Ophthalmic Biophysics Laboratory, Kallam Anji Reddy campus, L V Prasad Eye Institute, Hyderabad, India. Design: Laboratory study. Methods: Experiments were performed on IOLs from 2 different manufacturers (APPALENS 207, Appasamy Associates and SN60WF, Alcon Laboratories, Inc.). IOL powers ranged from 17 to 25 diopters (D). The IOLs were mounted in a fluid-filled chamber, and the on-axis and off-axis powers were measured using a laser ray tracing system over the central 3 mm zone with delivery angles ranging from −30 to +30 degrees in 5-degree increments. The position of the best focus was calculated for each IOL at each angle. The angular dependence of IOL power was compared with theoretical predictions. Results: Peripheral defocus increased significantly with increasing incidence angle and power. The peripheral defocus at ±30 degrees increased from 5.8 to 8.5 D when the power increased from 17.5 to 24.5 D for APPALENS 207 and from 4.9 to 7.4 D when the power increased from 17 to 25 D for SN60WF. The mean difference between the measured and theoretical tangential power at ±30 degrees was 0.50 ± 0.16 D for the APPALENS 207 and −0.40 ± 0.10 D for the SN60WF, independent of IOL power. Conclusions: IOLs introduce a significant amount of peripheral defocus which varies significantly with IOL power and design. Given that peripheral defocus is related to lens power, replacement of the crystalline lens (approximately 24 D) with an IOL will produce a significant difference in peripheral defocus profile after surgery.Keywords
This publication has 18 references indexed in Scilit:
- Peripheral visual field shifts after intraocular lens implantationJournal of Cataract & Refractive Surgery, 2023
- Age-Dependence of the Peripheral Defocus of the Isolated Human Crystalline LensInvestigative Ophthalmology & Visual Science, 2021
- Age-Related Changes to the Three-Dimensional Full Shape of the Isolated Human Crystalline LensInvestigative Ophthalmology & Visual Science, 2020
- Changes in Monkey Crystalline Lens Spherical Aberration During Simulated Accommodation in a Lens StretcherInvestigative Ophthalmology & Visual Science, 2015
- Peripheral Defocus and Myopia Progression in Myopic Children Randomly Assigned to Wear Single Vision and Progressive Addition LensesInvestigative Ophthalmology & Visual Science, 2013
- Comparison of the Optical Image Quality in the Periphery of Phakic and Pseudophakic EyesInvestigative Ophthalmology & Visual Science, 2013
- Peripheral refraction in pseudophakic eyes measured by infrared scanning photoretinoscopyJournal of Cataract & Refractive Surgery, 2012
- Refractive index measurement of the isolated crystalline lens using optical coherence tomographyVision Research, 2008
- Shape of the Retinal Surface in Emmetropia and MyopiaInvestigative Ophthalmology & Visual Science, 2005
- Peripheral Power Errors and Astigmatism of Eyes Corrected with Intraocular LensesOptometry and Vision Science, 1991