Directly measured mucosal pressures produced by the i‐gel TM and laryngeal mask airway Supreme TM in paralysed anaesthetised patients
Authors:
Open Access
- 11 February 2012
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 67 (4) , 407-410
Abstract
Summary: The i‐gel TM and LMA Supreme TM are extraglottic airway devices with non‐inflatable and inflatable cuffs, respectively. We hypothesised that directly measured mucosal pressures would differ between these devices in anesthetised paralysed patients. Thirty patients were randomly allocated to receive one of these two devices. Four pressure sensors were attached to all airway devices used to measure mucosal pressure at the base of the tongue, the distal oropharynx, the hypopharynx and the pyriform fossa. At these four places, median (IQR [range]) i‐gel mucosal pressures were 8.0 (2.7–10.7 [0–26.7]), 5.0 (2.7–7 [1.0–37.3]), 9.3 (2.7–13.3 [0–22.7] and 8.0 (2.7–10.7 [0–25.3]) cmH 2 O, respectively, and for the LMA Supreme, these were 5.0 (0.5–8.0 [0–33]), 4.0 (1.3–9.3 [0–24]), 10.7 (4–17.3 [0–26.7]) and 8.0 (0–10.7 [0–36]) cmH 2 O, respectively. Mucosal pressures were low and similar for both devices. The LMA Supreme mucosal pressures were higher in the hypopharynx than in the distal oropharynx (p = 0.04) and base of the tongue (p = 0.011). There were no pressure differences between the locations for the i‐gel.Keywords
This publication has 16 references indexed in Scilit:
- Crossover Comparison of the Laryngeal Mask Supreme™ and the i-gel™ in Simulated Difficult Airway Scenario in Anesthetized PatientsAnesthesiology, 2009
- A randomised crossover trial comparing the i‐gel supraglottic airway and classic laryngeal mask airway*Anaesthesia, 2009
- The Laryngeal Mask Airway Supreme TM – a single use laryngeal mask airway with an oesophageal vent. A randomised, cross‐over study with the Laryngeal Mask Airway ProSeal TM in paralysed, anaesthetised patientsAnaesthesia, 2008
- Measurement of tracheal wall pressure: a comparison of three different in vitro techniquesAnaesthesia, 2008
- Suction catheter guided insertion of the ProSeal™ laryngeal mask airway is superior to the digital techniqueCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2006
- Gum Elastic Bougie-guided Insertion of the ProSeal™ Laryngeal Mask Airway Is Superior to the Digital and Introducer Tool TechniquesAnesthesiology, 2004
- Pharyngeal Mucosal Pressures with the Laryngeal Tube Airway™ versus ProSeal Laryngeal Mask Airway™AINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie, 2003
- The Influence of Cuff Volume and Anatomic Location on Pharyngeal, Esophageal, and Tracheal Mucosal Pressures with the Esophageal Tracheal CombitubeAnesthesiology, 2002
- Pharyngeal Mucosal Pressure and PerfusionAnesthesiology, 1999
- Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patientsBritish Journal of Anaesthesia, 1999