The Effects of Additional Filtration on Image Quality and Radiation Dose in Cone Beam CT: An In Vivo Preliminary Investigation
Open Access
- 1 January 2022
- journal article
- research article
- Published by Wiley in BioMed Research International
- Vol. 2022 (1) , 7031269
Abstract
Purpose. The aim of this study was to investigate the effect of reduced radiation doses on the image quality of cone‐beam computed tomography scans and the suitability of such imaging for orthodontics, oral surgery, dental implantology, periodontology, and endodontology. Materials and Methods. Cone‐beam computed tomography scans of a live patient were performed using seven attenuation filters with increased thickness to decrease the effective radiation dose from 22.4 to 1.8 μSv, and the effects of different radiation doses on image quality were further analysed. Quantitative image quality was calculated using dedicated measures, such as signal and contrast‐to‐noise ratio and sharpness. A panel of five certified raters assessed the cone‐beam computed tomography scans qualitatively. Nine anatomical structures relevant to dentistry were identified, and the overall acceptance was assessed. Results. Linear reduction of the effective radiation dose had a nonlinear effect on image quality. A 5‐fold reduction in the effective dose led to acceptable quantitative and qualitative image quality measures, and the identification rate of dental anatomical structures was 80% or greater. The use of less than 40% of the reference dose was unacceptable for all dental specialties. Conclusions. The ideal radiation dose for specific diagnostic requirements remains a patient‐related and specialty‐related decision that must be made on an individual basis. Based on the results of this study, it is possible to reduce exposure in selected patients, and at the same time obtain sufficient quality of images for clinical purposes.Keywords
This publication has 41 references indexed in Scilit:
- A comprehensive study on the relationship between the image quality and imaging dose in low-dose cone beam CTPhysics in Medicine & Biology, 2012
- Comparative study between conventional and cone beam CT-synthesized half and total skull cephalogramsDentomaxillofacial Radiology, 2012
- Landmark identification errors on cone-beam computed tomography-derived cephalograms and conventional digital cephalogramsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2011
- A manufacturer's role in reducing the dose of cone beam computed tomography examinations: effect of beam filtrationDentomaxillofacial Radiology, 2011
- The current status of cone beam computed tomography imaging in orthodonticsDentomaxillofacial Radiology, 2011
- A comparison between 2D and 3D cephalometry on CBCT scans of human skullsInternational Journal of Oral and Maxillofacial Surgery, 2010
- Modulation transfer function evaluation of cone beam computed tomography for dental use with the oversampling methodDentomaxillofacial Radiology, 2010
- Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: A systematic review of the literatureInternational Journal of Oral and Maxillofacial Surgery, 2009
- In Vivo Comparison of Conventional and Cone Beam CT Synthesized CephalogramsThe Angle Orthodontist, 2008
- Comparison of conventional and cone beam CT synthesized cephalogramsDentomaxillofacial Radiology, 2007