Capacitive versus Overlap Decoupling of Adjacent Radio Frequency Phased Array Coil Elements: An Imaging Robustness Comparison When Sample Load Varies for 3 Tesla MRI
Open Access
- 15 December 2020
- journal article
- research article
- Published by Wiley in Concepts in Magnetic Resonance Part B
- Vol. 2020, 1-14
Abstract
Phased array (PA) receive coils are built such that coil elements approximate independent antenna behavior. One method of achieving this goal is to use an available decoupling method to decouple adjacent coil elements. The purpose of this work was to compare the relative performance of two decoupling methods as a function of variation in sample load. Two PA receive coils with 5 channels (5-ch) each, equal outer dimensions, and formed on 12 cm diameter cylindrical phantoms of conductivities 0.3, 0.6, and 0.9 S/m were evaluated for relative signal-to-noise ratio (SNR) and parallel imaging performance. They were only tuned and matched to the 0.6 S/m phantom. Simulated and measured axial, sagittal, and coronal 5-ch PA coil SNR ratios were compared by dividing the overlap by the capacitive decoupled coil SNR results. Issues related to the selection of capacitor values for the two decoupling methods were evaluated by taking the ratio of the match and tune capacitors for large and small 2 channel (2-ch) PA coils. The SNR ratios showed that the SNR of the two decoupling methods were very similar. The inverse geometry-factor maps showed similar but better overall parallel imaging performance for the capacitive decoupled method. The quotients for the 2-ch PA coils’ maximum and minimum capacitor value ratios are 3.28 and 1.38 for the large and 3.28 and 2.22 for the small PA. The results of this paper demonstrate that as the sample load varies, the capacitive and overlap decoupling methods are very similar in relative SNR and this similarity continues for parallel imaging performance. Although, for the 5-ch coils studied, the capacitive decoupling method has a slight SNR and parallel imaging advantage and it was noted that the capacitive decoupled coil is more likely to encounter unbuildable PA coil configurations.Keywords
Funding Information
- Siemens Healthcare (R01 HL 127582)
- National Institutes of Health (R01 HL 127582)
This publication has 32 references indexed in Scilit:
- Two‐Dimensional sixteen channel transmit/receive coil array for cardiac MRI at 7.0 T: Design, evaluation, and applicationJournal of Magnetic Resonance Imaging, 2012
- A numerical postprocessing procedure for analyzing radio frequency MRI coilsConcepts in Magnetic Resonance Part A, 2011
- Design and application of a four‐channel transmit/receive surface coil for functional cardiac imaging at 7TJournal of Magnetic Resonance Imaging, 2011
- An Eight-Channel Phased Array RF Coil for Spine MR Imaging at 7 TInvestigative Radiology, 2009
- Image reconstruction in SNR units: A general method for SNR measurement†Magnetic Resonance in Medicine, 2005
- Coupling and decoupling theory and its application to the MRI phased arrayMagnetic Resonance in Medicine, 2002
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- A 3 × 3 Mesh Two-Dimensional Ladder Network Resonator for MRI of the Human HeadJournal of Magnetic Resonance, Series B, 1995
- A 64 MHz half-birdcage resonator for clinical imagingJournal of Magnetic Resonance (1969), 1990
- Parallel image acquisition from noninteracting local coilsJournal of Magnetic Resonance (1969), 1986