Beam rate influence on dose distribution and fluence map in IMRT dynamic technique
- 13 February 2012
- journal article
- Published by VM Media Group sp. z o.o in Reports of Practical Oncology and Radiotherapy
- Vol. 17 (2) , 97-103
Abstract
To examine the impact of beam rate on dose distribution in IMRT plans and then to evaluate agreement of calculated and measured dose distributions for various beam rate values. Accelerators used in radiotherapy utilize some beam rate modes which can shorten irradiation time and thus reduce ability of patient movement during a treatment session. This aspect should be considered in high conformal dynamic techniques. Dose calculation was done for two different beam rates (100 MU/min and 600 MU/min) in an IMRT plan. For both, a comparison of Radiation Planning Index (RPI) and MU was conducted. Secondly, the comparison of optimal fluence maps and corresponding actual fluence maps was done. Next, actual fluence maps were measured and compared with the calculated ones. Gamma index was used for that assessment. Additionally, positions of each leaf of the MLC were controlled by home made software. Dose distribution obtained for lower beam rates was slightly better than for higher beam rates in terms of target coverage and risk structure protection. Lower numbers of MUs were achieved in 100 MU/min plans than in 600 MU/min plans. Actual fluence maps converted from optimal ones demonstrated more similarity in 100 MU/min plans. Better conformity of the measured maps to the calculated ones was obtained when a lower beam rate was applied. However, these differences were small. No correlation was found between quality of fluence map conversion and leaf motion accuracy. Execution of dynamic techniques is dependent on beam rate. However, these differences are minor. Analysis shows a slight superiority of a lower beam rate. It does not significantly affect treatment accuracy.Keywords
This publication has 20 references indexed in Scilit:
- Application of IMRT in adjuvant treatment of soft tissue sarcomas of the thigh—Preliminary resultsReports of Practical Oncology and Radiotherapy, 2011
- Comparison between intensity modulated radiotherapy (IMRT) and 3D tangential beams technique used in patients with early-stage breast cancer who received breast-conserving therapyReports of Practical Oncology and Radiotherapy, 2010
- Potentials of on-line repositioning based on implanted fiducial markers and electronic portal imaging in prostate cancer radiotherapyRadiation Oncology, 2009
- Impact of different leaf velocities and dose rates on the number of monitor units and the dose-volume-histograms using intensity modulated radiotherapy with sliding-window techniqueRadiation Oncology, 2008
- Radiation Planning Index for dose distribution evaluation in stereotactic radiotherapyReports of Practical Oncology and Radiotherapy, 2008
- Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in waterPhysics in Medicine & Biology, 2006
- QA of intensity-modulated beams using dynamic MLC log filesJournal of Medical Physics, 2006
- An extensive log‐file analysis of step‐and‐shoot intensity modulated radiation therapy segment delivery errorsMedical Physics, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Electronic Portal Imaging with On-Line Correction of Setup Error in Thoracic Irradiation: Clinical EvaluationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1998