    A Hand-Held 190+190 Row–Column Addressed CMUT Probe for Volumetric Imaging

    by Rune Sixten Grass, Mathias Engholm, Andreas Spandet Havreland, Christopher Beers, Martin Lind Ommen, Stine Lovholt Grue Pedersen, Lars Nordahl Moesner, Matthias Bo Stuart, Mudabbir Tufail Bhatti, Borislav G. Tomov, et al.

    This paper presents the design, fabrication, and characterization of a 190+190 row-column addressed (RCA) capacitive micromachined ultrasonic transducer (CMUT) array integrated in a custom hand-held probe handle. The array has a designed 4.5 MHz center frequency in immersion and a pitch of $95~\mu $ m which corresponds to $\approx ~\lambda $ /4. The array has a $2.14\times2.14$ cm $^{2}$ footprint including an integrated apodization scheme to reduce ghost echoes when performing ultrasound imaging. The array was fabricated using a combination of fusion and anodic bonding, and a deposit, remove, etch, multistep (DREM) etch to reduce substrate coupling and improve electrode conductivity. The transducer array was wire-bonded to a rigid-flex printed circuit board (PCB), encapsulated in room temperature vulcanizing (RTV) silicone polymer, electromagnetic interference (EMI) shielded, and mounted in a 3D-milled PPSU probe handle. The probe was characterized using the SARUS experimental scanner and 3D volumetric imaging was demonstrated on scatter and wire phantoms. The imaging depth was derived from tissue mimicking phantom measurements (0.5 dB MHz $^{-1} \text{cm}^{-1}$ attenuation) by estimating the SNR at varying depths. For a synthetic aperture imaging sequence with 96+96 emissions the imaging depth was 3.6 cm. The center frequency measured from the impulse response spectra in transmit and pulse-echo was 6.0 ± 0.9 MHz and 5.3 ± 0.4 MHz, and the corresponding relative bandwidths were 62.8 ± 4.5 % and 86.2 ± 10.4 %. The fabrication process showed clear improvement in relative receive sensitivity and transmit pressure uniformity compared to earlier silicon-on-insulator (SOI) based designs. However, at the same time it presented yield problems resulting in only around 55 % elements with a good response.


  • Imaging Performance for Two Row–Column Arrays

    by Hamed Bouzari, Mathias Engholm, Svetoslav Ivanov Nikolov, Matthias Bo Stuart, Erik Vilain Thomsen, Jorgen Arendt Jensen

    This study evaluates the volumetric imaging performance of two prototyped 62+62 row–column-addressed (RCA) 2-D array transducer probes using three Synthetic Aperture Imaging (SAI) emission sequences and two different beamsformers. The probes are fabricated using capacitive micromachined ultrasonic transducer (CMUT), and piezoelectric transducer (PZT) technology. Both have integrated apodization to reduce ghost echoes and are designed with similar acoustical features i.e., 3MHz center frequency, l/2-pitch, and 24.8×24.8mm2 active footprint. Raw RF data are obtained using an experimental research ultrasound scanner, SARUS. The SAI sequences are designed for imaging down to 14 cm at a volume rate of 88 Hz. Two beamforming methods: Spatial matched filtering and rowcolumn adapted delay-and-sum are used for beamforming the RF data. The imaging quality is investigated through simulations and phantom measurements. Both probes on average have similar lateral full-width at half-maximum (FWHM) values, but the PZT probe has 20% smaller cystic resolution values and 70% larger contrast-to-noise ratio compared to the CMUT probe. The CMUT probe can penetrate down to 15 cm, and the PZT probe down to 30 cm. The CMUT probe has 17% smaller axial FWHM values. The matched filter focusing shows and improved B-mode image for measurements on a cyst phantom with an improved speckle pattern and better visualization of deeper lying cysts. The results of this study demonstrate the potentials of RCA 2-D arrays against fully addressed 2-D arrays, which are low channel count (e.g. 124 instead of 3,844), low acoustic intensity (MI ≤0.88 and Ispta ≤5.5mW/cm2), and high penetration depth (down to 30 cm), which makes 3-D imaging at high volume rates possible with equipment in the price range of conventional 2-D imaging.


  • CMUT Electrode Resistance Design: Modeling and Experimental Verification by a Row-Column Array

    by Andreas Spandet Havreland, Mathias Engholm, Borislav Gueorguiev Tomov, Jorgen Arendt Jensen, Ole Hansen, Erik Vilain Thomsen

    This paper addresses the importance of having control over the resistivity of the electrodes for capacitive micromachined ultrasonic transducers (CMUT) devices. The electrode resistivity can vary depending on the fabrication technology used, and resistivity control becomes especially important in the cases where metal electrodes can not be used. This raises the question: When is the resistivity of an electrode sufficiently low? To answer this question we have developed a simple design criterion. The criterion describes the attenuation of AC signals along a CMUT element. It is shown that the non-dimensional product between angular excitation frequency, resistance, and capacitance ωRC of an element has to be smaller than 0.35 to ensure an AC potential drop along the element of less than 1%. The optimal magnitude and directionality of the transmit pressure will be achieved if CMUT elements are designed according to the developed criteria. Hence, the model can be used to estimate device parameters that will ensure the CMUT is suitable for generating ultrasound images. An example is given where the model is used to predict the required electrode thickness for structured electrodes made of Gold, Aluminium, and Indium-Tin-Oxide, respectively. To verify the model, two Row-Column addressed (RCA) CMUT transducers were used to illustrate the effect of high and low electrode resistivity. One transducer had a sufficient electrode resistivity, and the other had an insufficient electrode resistivity. The RCA CMUT transducers were fabricated using fusion bonding, where the top electrode is made of aluminium and the bottom electrode is made of doped silicon. The resistivity of the aluminium top electrode is 2×10-6 Ωcm for both transducers, whereas the resistivity for the bottom electrode is 0.1 Ωcm for the first transducer and 0.005 Ωcm for the second transducer. The transducer with low resistivity emits pressure uniformly along both the rows and columns, whereas the transmit pressure field from the other transducer has a uniformly distributed pressure field along the rows, but a decreasing pressure field along the columns due to the high resistivity in the bottom electrode. The pressure drop, along the columns is frequency dependent and has been observed to be 63%, 74%, and 82% for the excitation frequencies 2 MHz, 4.5 MHz, and 7 MHz, respectively.


  • Open Access

    Increasing the field-of-view of row–column-addressed ultrasound transducers: implementation of a diverging compound lens

    by Mathias Engholm, Christopher Beers, Hamed Bouzari, Jørgen Arendt Jensen, Erik Vilain Thomsen


  • Curvilinear 3-D Imaging Using Row–Column-Addressed 2-D Arrays With a Diverging Lens: Phantom Study

    by Hamed Bouzari, Mathias Engholm, Christopher Beers, Matthias Bo Stuart, Svetoslav Ivanov Nikolov, Erik Vilain Thomsen, Jorgen Arendt Jensen

    A double-curved diverging lens over the flat row-column-addressed (RCA) 2-D array can extend its inherent rectilinear 3-D imaging field of view (FOV) to a curvilinear volume region, which is necessary for applications such as abdominal and cardiac imaging. Two concave lenses with radii of 12.7 and 25.4 mm were manufactured using RTV664 silicone. The diverging properties of the lenses were evaluated based on simulations and measurements on several phantoms. The measured FOV for both lenses in contact with tissue mimicking phantom was less than 15% different from the theoretical predictions, i.e., a curvilinear FOV of 32° × 32° and 24°×24° for the 12.7and 25.4-mm radii lenses. A synthetic aperture imaging sequence with single-element transmissions was designed for imaging down to 140 mm at a volume rate of 88 Hz. The performance was evaluated in terms of signal-to-noise ratio, FOV, and full-width at half-maximum (FWHM) of a focused beam. The penetration depths in a tissue mimicking phantom with 0.5-dB/(cm·MHz) attenuation were 100 and 125 mm for the lenses with radii of 12.7 and 25.4 mm. The azimuth, elevation, and radial FWHM at 43-mm depth were (5.8, 5.8, 1)λ and (6, 6, 1)λ. The results of this study confirm that the proposed lens approach is an effective method for increasing the FOV, when imaging with RCA 2-D arrays.


  • Open Access

    Probe development of CMUT and PZT row–column-addressed 2-D arrays

    by Mathias Engholm, Hamed Bouzari, Thomas Lehrmann Christiansen, Christopher Beers, Jan Peter Bagge, Lars Nordahl Moesner, Søren Elmin Diederichsen, Matthias Bo Stuart, Jørgen Arendt Jensen, Erik Vilain Thomsen


  • Improved focusing method for 3-D imaging using row-column-addressed 2-D arrays

    by Hamed Bouzari, Mathias Engholm, Matthias Bo Stuart, Erik Vilain Thomsen, Jorgen Arendt Jensen

    A row-column-addressed (RCA) 2-D array can be interpreted as two orthogonal 1-D arrays. By transmitting with row elements and receiving the echoes through column elements or vice versa, a rectilinear volume in front of the array can be beamformed. Since the transmit and receive 1-D arrays are orthogonal to each other, only one-way focusing is possible in each transmit or receive plane. For applications, where the scatterers are sparse, e.g., in micro-bubble tracking, this study suggests to multiply the envelope data received by the row elements when transmitting with columns as well as the data received by the column elements when transmitting with rows, to improve the focusing. In this way, at each point a two-way focused profile in both transmit and receive directions can be produced. This paper investigates the performance of the new focusing scheme based on simulations and phantom measurements with a PZT λ/2-pitch 3 MHz 62+62 RCA 2-D transducer probe. A synthetic aperture imaging sequence with single element transmissions at a time, is designed for imaging down to 14 cm at a volume rate of 44 Hz.


  • Output pressure and pulse-echo characteristics of Capacitive Micromachined Ultrasonic Transducers as function of plate thickness

    by Soren Elmin Diederichsen, Jesper Mark Fly Hansen, Mathias Engholm, Jorgen Arendt Jensen, Erik Vilain Thomsen

    The energy transduction of a Capacitive Micromachined Ultrasonic Transducer (CMUT) depends on the movement of a flexible plate, which has a low mechanical impedance compared to its bulky piezoelectric counterpart. Consequently, the output pressure of a CMUT is generally lower. This limits the penetration depth, and thus the visualization of deeper structures in the body, as well as non-linear imaging such as tissue harmonic imaging. The objective of this work is to investigate how the output pressure and pulse-echo signal of a CMUT scales with the (silicon) plate thickness.


  • 3-D imaging using row-column-addressed 2-D arrays with a diverging lens: Phantom study

    by Hamed Bouzari, Mathias Engholm, Christopher Beers, Matthias Bo Stuart, Svetoslav Ivanov Nikolov, Erik Vilain Thomsen, Jorgen Arendt Jensen

    A double-curved diverging lens over a flat row-column-addressed (RCA) 2-D array can extend its inherent rectilinear 3-D imaging field-of-view (FOV) to a curvilinear volume region, which is necessary for applications such as abdominal and cardiac imaging. A concave lens with radius of 12.7 mm was manufactured using RTV664 silicone. The diverging properties of the lens were evaluated based on measurements on several phantoms. The measured 6 dB FOV in contact with a material similar to human soft tissue was less than 15% different from the theoretical predictions, i.e., a curvilinear FOV of 32°×32°. A synthetic aperture imaging sequence with single element transmissions was designed for imaging down to 14 cm at a volume rate of 88 Hz. The performance was evaluated in terms of signal-to-noise ratio (SNR), FOV, and full-width-at-half-maximum (FWHM). The penetration depth in a tissue mimicking phantom with 0.5 dB/(cm MHz) attenuation was 13 cm. The results of this study confirm that the proposed lens approach is an effective method for increasing the FOV, when imaging with RCA 2-D arrays.


  • Output pressure and pulse-echo characteristics of CMUTs as function of plate dimensions

    by Soren Elmin Diederichsen, Jesper Mark Fly Hansen, Mathias Engholm, Jorgen Arendt Jensen, Erik Vilain Thomsen

    This paper presents an experimental study of the acoustic performance of Capacitive Micromachined Ultrasonic Transducers (CMUTs) as function of plate dimensions. The objective is to increase the output pressure without decreasing the pulse-echo signal. The CMUTs are fabricated with a LOCOS process, followed by direct wafer fusion bonding to a Silicon-On-Insulator (SOI) wafer. In this way, the plate thickness is determined by the SOI wafer device layer thickness, resulting in CMUTs with plate thicknesses of 2, 9.3 and 15 μm. The corresponding radii and gap heights resulting in an immersion frequency of 5MHz and a pull-in voltage of 200V are obtained using finite element analysis. Hydrophone and plane reflector measurements are used to assess the acoustic performance. Increasing the plate thickness from 2μm to 15μm decreases the pulse-echo bandwidth from >100% to 30%. A maximum in both peak-to-peak output pressure and pulse-echo signal is obtained for the 9.3μm plate, which still has a moderate pulseecho bandwidth of 60%. The 9.3μm plate results in a 1.9 times higher peak-to-peak output pressure and a 3.6 times higher pulse-echo signal compared to the 2μm plate. By adjusting the plate dimensions of a CMUT it is possible to optimize its acoustic performance for medical imaging applications, including visualization of deeper structures in the body, as well as nonlinear imaging such as tissue harmonic imaging.