Abstract
A soft x-ray diffraction grating with an aperiodic multilayer coating was designed to enhance spectral flux in a photon energy range of 1–2 keV by means of numerical calculation. Optimized groove depth and angle of incidence were 2.65 nm and 87.50°, respectively, for the grating with a groove density of 2400 lines/mm. The optimum number of W/B4C layer pairs was 6; the thicknesses of four bottom B4C layers were constant, and those of two top layers were unequal, while all W layer thicknesses were the same. An additional top layer of Au was crucial to provide uniform diffraction efficiency across the spectral range. In the region of 1–2 keV, the calculated diffraction efficiency of the designed gratings was ∼4.4% on average, and the calculated spectral flux (which correlates with analytical sensitivity) was 2.8 times larger than that of a previously designed Au-coated grating with an angle of incidence of 88.65°. The designed grating was manufactured; its measured diffraction efficiency was close to the calculated one in the 0.6–1.4 keV and >2 keV spectral ranges but somewhat lower in the 1.6–1.8 keV range. The spectrograph equipped with the new grating exhibited resolving power exceeding 400, which approached the pixel-size limit.
Funding Information
  • Japan Society for the Promotion of Science (JP19H00669)
  • Japan Society for the Promotion of Science (JP19K12642)
  • Japan Society for the Promotion of Science (JP25H00621)

This publication has 21 references indexed in Scilit: