Abstract
We report on the study of a microwave superconducting quantum interference device (SQUID) multiplexer (MW-Mux) with current noise lower than that of our typical transition-edge sensor (TES) X-ray microcalorimeter. A large mutual inductance, between SQUID loop and input inductor connected to a TES X-ray microcalorimeter in series, realizes such low current noise referred to the SQUID input, but also can cause undesirable resonances which affect the SQUID performances. In this paper, we developed the MW-Mux with mutual inductance of about 370 pH, which is six times larger than that of the previous one, by adopting a spiral input coil longer than a half wavelength at the readout microwave frequency. Its characteristics were evaluated, and the results showed that the distortion of the SQUID characteristics and the degradation of its noise performance were not clearly observed. Finally, we have developed a MW-Mux microstrip SQUID with a low current noise of $\rm \lt 20 pA/\sqrt{Hz}$ which is below the typical noise of our TES X-ray microcalorimeter, $\rm 30 pA/\sqrt{Hz}$ .
Funding Information
  • JSPS KAKENHI (18H01260, 2622070)

This publication has 17 references indexed in Scilit: