Initial Evaluation of the Performance of Novel Inorganic Scintillating Detectors for Small Animal Irradiation Dosimetry
- 8 January 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 20 (9) , 4704-4712
Abstract
The purpose of this study was to design and evaluate the performance of four novel inorganic scintillating detectors (ISDs) on the Small Animal Radiation Research Platform (SARRP). Relative scintillator output, measurement repeatability, setup uncertainty, linearity with dose rate, and signal reproducibility over time were investigated. The Gd 2 O 2 S:Tb detector had the highest relative signal output, generating up to 219 times more charge than a previously characterized BCF-60-based plastic scintillating detector (PSD). The Gd 2 O 2 S:Tb detector was then used to measure 220 kVp therapy beam profiles of 10 x 10 and 5 x 5 mm 2 fields. Beam profiles using the ZnS-based phosphor were also obtained and compared to investigate the performance of a lower density inorganic scintillator. 10 x 10 and 5 x 5 mm 2 therapy beam profile measurements made with the Gd 2 O 2 S:Tb and BCF-60 detectors differed, on average, by 1.1% and 1.9%, respectively. The ZnS:Ag measurements differed, on average, by 2.5% and 6% relative to BCF-60 measurements of the 10 x 10 and 5 x 5 mm 2 beam profiles, respectively. MicroCT imaging of the detector volumes was also performed, revealing poor packing of the ZnS:Ag crystalline phosphor in the deepest region of the cylindrical cavity. The Gd 2 O 2 S:Tb detector, in particular, has proven to be a promising candidate for real-time dosimetry of small fields in small animal irradiators, primarily because of the very large signal intensities observed, along with good repeatability, dose rate linearity, reproducibility and agreement with beam profile measurements made with a previously validated detector.Keywords
Funding Information
- Irish Association of Physicists in Medicine through the 2019 Young Investigator Grant
This publication has 25 references indexed in Scilit:
- On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetryPhysics in Medicine & Biology, 2013
- Characterization of image quality and image‐guidance performance of a preclinical microirradiatorMedical Physics, 2011
- High-Resolution, Small Animal Radiation Research Platform With X-Ray Tomographic Guidance CapabilitiesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Scintillation detectors for x-raysMeasurement Science and Technology, 2006
- Measurement accuracy and Cerenkov removal for high performance, high spatial resolution scintillation dosimetryMedical Physics, 2005
- Spectral discrimination of Čerenkov radiation in scintillating dosimetersMedical Physics, 2005
- Inorganic scintillators in medical imaging detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Experimental correction for the inner-filter effect in fluorescence spectraThe Analyst, 1994
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Visible Radiation Produced by Electrons Moving in a Medium with Velocities Exceeding that of LightPhysical Review B, 1937