Time‐resolved mapping of myocardial stiffness using 2D multifrequency spiral MR elastography with and without external vibration
Authors:
- Matthias S. AndersMAMatthias S. Anders
- Carsten WarmuthCWCarsten Warmuth
- Tom MeyerTMTom Meyer
- Helge HerthumHHHelge Herthum
- Mehrgan ShahryariMSMehrgan Shahryari
- Jakob SchattenfrohJSJakob Schattenfroh
- Corona MetzCMCorona Metz
- Jan BielingJBJan Bieling
- Josef PfeufferJPJosef Pfeuffer
- Simon VeldhoenSVSimon Veldhoen
- Jeanette Schulz‐MengerJSJeanette Schulz‐Menger
- Tobias SchaeffterTSTobias Schaeffter
Open Access
- 28 July 2025
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 94 (6) , 2421-2434
Abstract
Purpose: There is a clinical need for stiffness mapping of the heart; however, current cardiac magnetic resonance elastography (cMRE) has limited spatiotemporal resolution. Therefore, we developed 2D spiral multifrequency MRE of the human heart and conducted a study to analyze the consistency and reproducibility of motion‐encoding and stiffness mapping with and without external vibration.Methods: Eleven healthy volunteers were studied using single‐slice gradient‐echo spiral cMRE with cardiac triggering and encoding of harmonic shear wave fields at 70, 80, and 90 Hz frequency generated by either external drivers or endogenous heart motion. Tissue displacement was monitored synchronized to the cardiac cycle, and frequency‐resolved shear wave speed (SWS) maps were reconstructed as a proxy for left ventricular (LV) stiffness variations. After several days, all subjects underwent repeat scanning for reproducibility analysis based on intraclass correlation coefficients (ICCs).Results: cMRE with external vibration showed LV SWS to be highest in end‐systole (ES) (2.17 ± 0.23 m/s), followed by diastole (DIA) (1.94 ± 0.15 m/s) and isovolumetric contraction (IVC) (1.78 ± 0.18 m/s). ICCs decreased with distance from the R‐wave from excellent (0.93) in IVC to moderate (0.68) in ES. Without external vibration, sufficient LV harmonic displacement amplitudes permitted SWS reconstruction, resulting in similar SWS, but lower ICC values than with external vibration in IVC and ES.Conclusions: Multifrequency cMRE offers high spatiotemporal resolution and reproducibility with external vibration. In addition, the technique allows the encoding of endogenous shear waves during cardiac phases with pronounced wall motion.Keywords
This publication has 51 references indexed in Scilit:
- Reproducibility of Cardiac Multifrequency MR Elastography in Assessing Left Ventricular Stiffness and ViscosityJournal of Magnetic Resonance Imaging, 2024
- Reduction of breathing artifacts in multifrequency magnetic resonance elastography of the abdomenMagnetic Resonance in Medicine, 2020
- Cardiac MR elastography for quantitative assessment of elevated myocardial stiffness in cardiac amyloidosisJournal of Magnetic Resonance Imaging, 2017
- In vivo quantification of myocardial stiffness in hypertensive porcine hearts using MR elastographyJournal of Magnetic Resonance Imaging, 2016
- In vivo, high‐frequency three‐dimensional cardiac MR elastography: Feasibility in normal volunteersMagnetic Resonance in Medicine, 2016
- In vivo wideband multifrequency MR elastography of the human brain and liverMagnetic Resonance in Medicine, 2015
- Measuring age‐dependent myocardial stiffness across the cardiac cycle using MR elastography: A reproducibility studyMagnetic Resonance in Medicine, 2015
- Magnetic resonance elastography as a method to estimate myocardial contractilityJournal of Magnetic Resonance Imaging, 2012
- Estimation of k‐space trajectories in spiral MRIMagnetic Resonance in Medicine, 2009
- Fractional encoding of harmonic motions in MR elastographyMagnetic Resonance in Medicine, 2007