Automatic image quality evaluation in digital radiography using for‐processing and for‐presentation images
Open Access
- 5 February 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (4) , e14285
Abstract
Purpose: To investigate the impact of digital image post‐processing algorithms on various image quality (IQ) metrics of radiographic images under different exposure conditions.Methods: A custom‐made phantom constructed according to the instructions given in the IAEA Human Health Series No.39 publication was used, along with the respective software that automatically calculates various IQ metrics. Images with various exposure parameters were acquired with a digital radiography unit, which for each acquisition produces two images: one for‐processing (raw) and one for‐presentation (clinical). Various examination protocols were used, which incorporate diverse post‐processing algorithms. The IQ metrics’ values (IQ‐scores) obtained were analyzed to investigate the effects of increasing incident air kerma (IAK) on the image receptor, tube potential (kVp), additional filtration, and examination protocol on image quality, and the differences between image type (raw or clinical).Results: The IQ‐scores were consistent for repeated identical exposures for both raw and clinical images. The effect that changes in exposure parameters and examination protocol had on IQ‐scores were different depending on the IQ metric and image type. The expected positive effect that increasing IAK and decreasing tube potential should have on IQ was clearly exhibited in two IQ metrics only, the signal difference‐to‐noise‐ratio (SDNR) and the detectability index (d’), for both image types. No effect of additional filtration on any of the IQ metrics was detected on images of either type. An interesting finding of the study was that for all different image acquisition selections the d’ scores were larger in raw images, whereas the other IQ metrics were larger in clinical images for most of the cases.Conclusions: Since IQ‐scores of raw and their respective clinical images may be largely different, the same type of image should be consistently used for monitoring IQ constancy and when results from different X‐ray systems are compared.Keywords
This publication has 11 references indexed in Scilit:
- The IAEA remote and automated quality control methodology for radiography and mammographyJournal of Applied Clinical Medical Physics, 2021
- Evaluation of image quality and patient exposure in fluoroscopy using a phantom: Is there any clinical relevance?European Journal of Radiology, 2021
- Correlation between physical measurements and observer evaluations of image quality in digital chest radiographyMedical Physics, 2020
- Performance Evaluation of Computed Tomography Systems - The Report of AAPM Task Group 2332019
- The impact of image processing algorithms on digital radiography of patients with metalic hip implantsPhysica Medica, 2019
- An investigation into the validity of utilising the CDRAD 2.0 phantom for optimisation studies in digital radiographyBritish Journal of Radiology, 2018
- Correlation of Contrast-Detail Analysis and Clinical Image Quality Assessment in Chest Radiography with a Human Cadaver StudyRadiology, 2012
- A method for calculating the dose length product from CT DICOM imagesBritish Journal of Radiology, 2011
- Correlation of image quality with exposure index and processing protocol in a computed radiography systemRadiation Protection Dosimetry, 2007
- Evaluation of a software package for automated quality assessment of contrast detail images—comparison with subjective visual assessmentPhysics in Medicine & Biology, 2005