Evaluation of the Presence of Biofilms in Corrosive Points in Surgical Instruments after Reprocessing
Authors:
Open Access
Abstract
Corrosive surgical instruments are routinely observed in central sterile services departments around the world. In addition to other risks, they can harbor microorganisms in the form of biofilms. Thus, this study aimed to evaluate whether biofilms intentionally formed at corrosion points on surgical instruments are removable by manual and automated cleaning followed by sterilization. Laboratory experiments were performed where samples of corroded surgical instruments in use in practice were evaluated for biofilm presence using a scanning electron microscope. No biofilms were observed in the samples subjected to manual and automated cleaning, nor in the samples in which there was no intentional biofilm formation. Residual organic matter without the presence of microorganisms was observed.Keywords
This publication has 8 references indexed in Scilit:
- Effectiveness of manual versus automated cleaning on Staphylococcus epidermidis biofilm removal from the surface of surgical instrumentsAmerican Journal of Infection Control, 2020
- Complex design of surgical instruments as barrier for cleaning effectiveness, favouring biofilm formationJournal of Hospital Infection, 2019
- Bacterial biofilm formation on implantable devices and approaches to its treatment and preventionHeliyon, 2018
- Outbreak of Pseudomonas aeruginosa Surgical Site Infections after Arthroscopic Procedures: Texas, 2009Infection Control & Hospital Epidemiology, 2011
- Instrumentation-Specific Infection After Anterior Cruciate Ligament ReconstructionSports Health: A Multidisciplinary Approach, 2009
- Is biofilm accumulation on endoscope tubing a contributor to the failure of cleaning and decontamination?Journal of Hospital Infection, 2004
- Septic Arthritis Following Arthroscopic Meniscus Repair: A Cluster of Three CasesArthroscopy, 1999
- How Bacteria StickScientific American, 1978