Fabrication, testing, and clinical validation of an epoxy resin-based range shifter for pencil beam scanning proton therapy
Authors:
- Kantaram DarekarKDKantaram Darekar
- Bhushankumar J PatilBPBhushankumar J Patil
- Akshay BhavkeABAkshay Bhavke
- Umesh GayakeUGUmesh Gayake
- Vysakh RaveendranVRVysakh Raveendran
- Rohidas PundeRPRohidas Punde
- Prakash ShindePSPrakash Shinde
- Sanjay D DholeSDSanjay D Dhole
- Chithambara Prabu ArunasalamCAChithambara Prabu Arunasalam
- Lalit ChaudhariLCLalit Chaudhari
- Siddhartha LaskarSLSiddhartha Laskar
- 1 October 2026
- journal article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 12 (5) , 055076
Abstract
Background and Objectives. Range shifters (RS) are used to treat shallow targets in pencil beam scanning (PBS) proton therapy. This study aims to validate an in-house-fabricated epoxy resin RS through mechanical and dosimetric measurements, evaluate the Monte Carlo (MC) algorithm dose‐calculation accuracy of the treatment planning system (TPS) beam model incorporating this device, and assess its clinical feasibility for PBS proton therapy applications. Methods. The RS was fabricated using a silicone mould and machined to the required dimensions after curing. It was further evaluated through material characterization, mechanical testing, and dosimetric validation. Material characterization included determining the elemental composition, mass density, and mean excitation energy, which were incorporated into the beam modelling. Mechanical validation comprised dimensional verification, fitment testing, and assessment of material integrity and uniformity. Dosimetric validation was performed by measuring water-equivalent thickness (WET), range parameters, spot size, absolute dose, and fluence, and comparing them with TPS-MC calculations. Results. The cured epoxy resin exhibited normalized elemental composition of H:8.0%, C:65.0%, N:0.1%, and O:26.9%, with mass density 1.15 g cm −3 and mean excitation energy 72.5 eV. The fabricated RS matched the reference design within ±1 mm and demonstrated satisfactory fitment, reproducibility, material integrity and uniformity. The measured WET agreed with TPS estimated values within 0.06 g cm −2 , while differences in range parameters were Conclusion. This study presents the first comprehensive clinical validation and evaluation of cured solid epoxy resin as an RS material for PBS proton therapy applications. The in-house-fabricated, cured epoxy resin-based RS demonstrated excellent agreement with TPS-MC calculations and experimental measurements, establishing its clinical feasibility.Keywords
Funding Information
- Advanced Centre for Treatment, Research and Education in Cancer (IEC9009333)
This publication has 27 references indexed in Scilit:
- Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithmsJournal of Applied Clinical Medical Physics, 2025
- New energy layer reduction method for rapid intensity‐modulated proton therapy (IMPT) using a ridge filterMedical Physics, 2025
- The use of a mini‐ridge filter with cyclotron‐based pencil beam scanning proton therapyMedical Physics, 2023
- Impact of air gap, range shifter, and delivery technique on skin dose in proton therapyMedical Physics, 2020
- Clinical commissioning of intensity‐modulated proton therapy systems: Report of AAPM Task Group 185Medical Physics, 2020
- Determination of surface dose in pencil beam scanning proton therapyMedical Physics, 2020
- Measurements of in-air spot size of pencil proton beam for various air gaps in conjunction with a range shifter on a ProteusPLUS PBS dedicated machine and comparison to the proton dose calculation algorithmsAustralasian Physics & Engineering Sciences in Medicine, 2019
- AAPM task group 224: Comprehensive proton therapy machine quality assuranceMedical Physics, 2019
- Impact of spot size variations on dose in scanned proton beam therapyPhysica Medica, 2019
- Quantifying the effect of air gap, depth, and range shifter thickness on TPS dosimetric accuracy in superficial PBS proton therapyJournal of Applied Clinical Medical Physics, 2017