Patient dose reduction for a localizer radiograph with an additional tin filter in chest–abdomen–pelvis, spine, and head computed tomography examinations
- 9 February 2023
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 16 (2) , 1-8
Abstract
This study aimed to evaluate the dose reduction potential of adding a tin filter to localizer radiographs (LR) on computed tomography (CT) examinations in both phantom and clinical studies. LRs were performed using combinations of 120 kVp and 20 mA (120/20), 100 kVp with a tin filter, and 50 mA or 20 mA (Sn100/50, Sn100/20). For the phantom experiment, entrance surface doses (ESD) of the LRs were evaluated for each protocol using an anthropomorphic phantom. This retrospective clinical study included 700 patients (300 for chest–pelvis, 200 for spine, and 200 for head CTs). The volume CT dose indices (CTDIvols) of the main CT scans were recorded and placed into one of three groups based on body mass index (BMI): underweight, normal-weight, and overweight, to evaluate the effect of LR acquisition conditions on the performance of the automatic tube current modulation technique of subsequent CT scans. The ESDs of all LRs with the Sn100/50 protocol were 0.03 mGy, a decrease of more than 80% compared to those of the 120/20 protocol. Moreover, the Sn100/20 protocol reduced ESD to 0.02 mGy. In chest–pelvis CT, there were no significant differences in the CTDIvol between with and without a tin filter for each BMI group. However, the lateral LRs with the tin filter on the spine CT slightly reduced the CTDIvol in normal-weight and overweight patients. Although there is room to optimize the acquisition conditions for larger patients, an additional tin filter for LR is a useful means to efficiently reduce ESDs.Keywords
This publication has 15 references indexed in Scilit:
- Entrance surface dose measurements using a small OSL dosimeter with a computed tomography scanner having 320 rows of detectorsRadiological Physics and Technology, 2016
- INFLUENCE OF LOCALIZER AND SCAN DIRECTION ON THE DOSE-REDUCING EFFECT OF AUTOMATIC TUBE CURRENT MODULATION IN COMPUTED TOMOGRAPHYRadiation Protection Dosimetry, 2016
- Low-Dose CT of the Paranasal Sinuses: Minimizing X-Ray Exposure with Spectral ShapingEuropean Radiology, 2016
- Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVpEuropean Journal of Radiology, 2015
- Ultralow-Dose Chest Computed Tomography for Pulmonary Nodule DetectionInvestigative Radiology, 2014
- Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom studyEuropean Radiology, 2014
- Assessment of patient dose from CT localizer radiographsMedical Physics, 2013
- Effective Doses from Scan Projection Radiographs of the Head: Impact of Different Scanning Practices and Comparison with Conventional RadiographyAmerican Journal of Neuroradiology, 2008
- Automatic exposure control in multichannel CT with tube current modulation to achieve a constant level of image noise: Experimental assessment on pediatric phantomsMedical Physics, 2007
- Patient Dose Reduction in CT Examinations by Optimising Scanogram AcquisitionRadiation Protection Dosimetry, 2001