Accuracy of thick and thin intraocular lens power formulas using paraxial vergence calculation
- 1 March 2025
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 51 (3) , 182-187
Abstract
Purpose: To compare the prediction errors (PEs) of several thick intraocular lens (IOL) formulas with a thin lens approach using variations of the same paraxial vergence calculation formula. Setting: Department of Ophthalmology, Randers Regional Hospital, Denmark. Design: Prospective, noninterventional study. Methods: Optical low coherence reflectometry biometry was prospectively and consecutively performed in 132 eyes with subsequent phacoemulsification and insertion of the same aspheric IOL model. Clinical refraction was performed 2 months postoperatively. Retrospectively, the same paraxial vergence formula was used and only the methods were varied for calculating the postoperative IOL position, thickness, and curvatures to construct 4 formulas: Næser I formula based on thick lens calculation using the manufacturer's cutting card information, Næser II formula based on thick lens calculation using calculated IOL data from open sources, Næser III formula based on thick lens calculation and a fixed IOL thickness of 0.62 mm, and Næser IV formula based on thin lens calculation with fixed IOL position 0.31 mm anterior to the postoperative capsule. Each formula was optimized separately, hereby assuring a mean error of zero. The PE was defined as the difference between the measured and predicted spherical equivalent refraction. Results: Mean absolute error amounted to 0.30 ± 0.26 diopters for all 4 formulas with no statistically significant difference. PE averaged zero for short, normal, and long eyes. Conclusions: The additional optical information provided by thick IOL calculations seems insignificant compared with other sources of error, related to the accuracy of IOL manufacturing, crystalline lens refractive index, and postoperative refraction.Keywords
This publication has 30 references indexed in Scilit:
- Considerations of a thick lens formula for intraocular lens power calculationZeitschrift für Medizinische Physik, 2022
- Accuracy of thick-lens intraocular lens power calculation based on cutting-card or calculated data for lens architectureJournal of Cataract & Refractive Surgery, 2019
- A comparison of two methods to calculate axial lengthJournal of Cataract & Refractive Surgery, 2019
- C constant: New concept for ray tracing–assisted intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2014
- Intraocular lens calculation accuracy limits in normal eyesJournal of Cataract & Refractive Surgery, 2008
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 2008
- Prediction of the effective postoperative (intraocular lens) anterior chamber depthJournal of Cataract & Refractive Surgery, 2006
- Predicting postoperative intraocular lens position and refractionJournal of Cataract & Refractive Surgery, 2004
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Calculation of the thickness of an intraocular lensJournal of Cataract & Refractive Surgery, 1993