Cumulative Transfer Function for Assessment of MRI-Induced RF Heating Risk in Pediatric Patients Implanted with Bifurcated Leads
Preprint
Abstract
Purpose: RF-induced heating remains a major barrier to MRI access for patients with epicardial cardiac implantable electronic devices (CIEDs). Although ISO/TS 10974 Tier-3 transfer function (TF) methods are established for unbranched leads, no analogous framework exists for bifurcated leads, in which branch asymmetry and inter-branch coupling may substantially alter heating. We developed and validated a cumulative transfer function (cTF) framework to address this gap. Methods: Following ISO/TS 10974 Tier-3 formalism, we measured, calibrated, and validated cTFs for a commercial 35 cm bipolar epicardial lead at 1.5 T. The framework explicitly accounts for branch-specific response and cross-branch coupling. Validation was performed with 24 canonical lead configurations in a homogeneous phantom and, without recalibration, in a heterogeneous anthropomorphic pediatric phantom with clinically derived trajectories. A single-branch TF approximation served as a comparator. The validated cTF was applied to predict RF heating across adult and pediatric human models at multiple imaging landmarks. Results: Compared with the single-branch TF approximation, the cTF reduced prediction error by nearly 70% in the primary validation dataset. In secondary validation, the cTF maintained low error across clinically relevant trajectories and imaging landmarks. In human models, the framework revealed marked anatomy- and landmark-dependent variation in predicted heating for the tested 35 cm lead, with low predicted heating in pediatric models and substantially higher heating in selected adult chest and upper abdominal imaging scenarios. Conclusion: The cTF provides a validated framework for RF-heating assessment of bifurcated leads and substantially improves prediction accuracy over single-branch TF approximations that neglect branch coupling.Keywords
Funding Information
- National Institutes of Health (R01EB034377)
- National Institutes of Health (R01HL168859)
- National Institutes of Health (T32EB025766)
This publication has 30 references indexed in Scilit:
- RF‐induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRIMagnetic Resonance in Medicine, 2023
- Magnetic resonance conditionality of abandoned leads from active implantable medical devices at 1.5 TMagnetic Resonance in Medicine, 2021
- Effect of Device Configuration and Patient's Body Composition on the RF Heating and Nonsusceptibility Artifact of Deep Brain Stimulation Implants During MRI at 1.5T and 3TJournal of Magnetic Resonance Imaging, 2020
- MRI‐Related Heating of Implants and Devices: A ReviewJournal of Magnetic Resonance Imaging, 2020
- Coupled transfer function model for the evaluation of implanted cables safety in MRIMagnetic Resonance in Medicine, 2020
- Experimental setup for transfer function measurement to assess RF heating of medical leads in MRI: Validation in the case of a single wireMagnetic Resonance in Medicine, 2017
- MRI‐based transfer function determination for the assessment of implant safetyMagnetic Resonance in Medicine, 2017
- Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loopsMagnetic Resonance in Medicine, 2016
- Measuring RF‐induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leadsMagnetic Resonance in Medicine, 2009
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007