Accuracy of a helium‐beam radiography system based on thin pixel detectors for an anthropomorphic head phantom

Authors:

Abstract
Background: Ion‐beam radiography is a promising technique to verify the range of ion‐beam radiotherapy treatments regularly. To detect and quantify the water‐equivalent thickness (WET) of potential anatomical changes, ion‐beam radiographs must provide a sufficient WET accuracy on the level of 1%. Purpose: In this work, we show an energy‐painted helium‐beam radiograph of an anthropomorphic head phantom acquired with thin silicon pixel detectors for the first time. Furthermore, we determine the WET accuracy of our helium‐beam radiography system for the especially heterogeneous skull base region, which is highly relevant for the treatment of head and neck and skull base tumors. Methods: With a detection system based on pixelated semiconducting Timepix detectors, we track single ions upstream and downstream of the head phantom. Furthermore, we measure their energy deposition in a thin Timepix detector behind the anthropomorphic phantom. To ensure a high precision of the image, we acquired a radiograph by using helium beams with five initial energies between 146.84 and 188.07 MeV/u following the energy painting algorithm. With a Siemens SOMATOM Confidence CT scanner, a single‐ and dual‐energy CT were acquired with clinical protocols and translated to relative stopping power (RSP) values. After projecting these scans, the resulting WET maps were compared to the helium‐beam radiograph. To evaluate the accuracy of all three modalities, a reference data set based on range‐pullback measurements and a segmentation of a high‐resolution CT scan was taken into account. Results: The mean absolute percentage error (MAPE) of all modalities was determined to be between 0.95% and 1.16%. Also, the root‐mean‐square percentage error (RMSPE) was similar for all modalities ranging from 1.19% to 1.46%. These deviations from the reference scan were found to mainly stem from an overestimation of air and sinus tissue and underestimation of cortical bone. Conclusions: The helium‐beam radiograph was shown to achieve a WET accuracy competitive with that of clinically used imaging methods. If certain technical aspects are addressed, helium‐beam radiography may emerge as an auspicious imaging modality for on‐couch range verification of ion‐beam radiotherapy treatments allowing for regular detection and quantification of anatomical changes.
Funding Information
  • Deutsche Forschungsgemeinschaft (426970603)

This publication has 55 references indexed in Scilit: