Cardiac output measurement with transpulmonary ultrasound dilution is feasible in the presence of a left-to-right shunt: a validation study in lambs
Open Access
- 14 December 2011
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 108 (3) , 409-416
Abstract
Cardiac output (CO) monitoring remains complex in newborns as most of the current technologies fail to accurately measure systemic blood flow in the presence of shunts. We validated CO measurements using transpulmonary ultrasound dilution (TPUD) in a neonatal lamb model with a left-to-right shunt. Regular arterial and central venous catheters were inserted into seven lambs (3.5–8.3 kg). A surgically constructed left-to-right aorto-pulmonary Gore-Tex® shunt was intermittently opened and closed, while CO was manipulated by creating haemorrhagic hypotension. CO measurements with TPUD (COtpud) were compared with those obtained by an ultrasonic transit-time flow probe positioned around the main pulmonary artery (COufp). We performed 72 sessions of three paired CO measurements. The mean COufp was 1.00 litre min−1 (range 0.47–1.75 litre min−1) and mean COtpud 1.05 litre min−1 (range 0.54–1.87 litre min−1). With an open shunt, the mean Qp/Qs ratio was 1.8 (range 1.3–2.6). A comparison between COufp and COtpud showed a mean bias (sd) of 0.03 (0.09) and 0.07 (0.10) litre min−1, respectively, for measurements with a closed and an open shunt. The percentage error was 18% and 20% for measurements with a closed and an open shunt. Polar plot analysis showed good trending ability for both closed and open shunt groups. TPUD is a reliable technology to measure CO in the presence of a left-to-right shunt.Keywords
This publication has 42 references indexed in Scilit:
- Validation of Cardiac Output Measurement by Ultrasound Dilution Technique With Pulmonary Artery Thermodilution in a Pediatric Animal ModelPediatric Cardiology, 2011
- A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac OutputAnesthesia & Analgesia, 2010
- Application of ultrasound dilution technology for cardiac output measurement: Cerebral and systemic hemodynamic consequences in a juvenile animal modelPediatric Critical Care Medicine, 2010
- Clinical practice parameters for hemodynamic support of pediatric and neonatal septic shock: 2007 update from the American College of Critical Care Medicine*Critical Care Medicine, 2009
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients*Pediatric Critical Care Medicine, 2008
- A NOVEL METHOD TO MEASURE CARDIAC OUTPUT IN THE PEDIATRIC ICU: ANIMAL VALIDATION AND PRELIMINARY CLINICAL STUDY.Critical Care Medicine, 2006
- Arterial pulse wave analysis: An accurate means of determining cardiac output in children*Pediatric Critical Care Medicine, 2006
- Low superior vena cava flow and neurodevelopment at 3 years in very preterm infantsThe Journal of Pediatrics, 2004