Impact of MRI RF coil design on the RF-induced heating of medical implants: fixed B 1 + rms exposure versus normal operating mode
Open Access
- 26 February 2024
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 69 (5) , 055021
Abstract
A direct comparison of the impact of RF coil design under specific absorption rate and B 1 + rms limitations are investigated and quantified using RF coils of different geometries and topologies at 64 MHz and 128 MHz. The RF-induced in vivo electric field and power deposition of a 50 cm long pacemaker and 55 cm long deep brain stimulator (DBS) are evaluated within two anatomical models exposed with these RF coils. The associated uncertainty is quantified and analyzed under a fixed B 1 + rms incident and normal operating mode. For a fixed B 1 + rms incident, the in vivo incident field shows a much higher uncertainty (>5.6 dB) to the RF coil diameter compared to other design parameters (e.g. B 1 + rms field to the untested implant will lead to a large variation in the induced incident and power deposition of the implant, as a result, a larger safe margin when different coil designs (e.g. coil diameter) are considered.Keywords
Funding Information
- Lanzhou City Youth Science and Technology Talent Innovation Project (2023-QN-87)
- National Natural Science Foundation of China (62101226)
This publication has 16 references indexed in Scilit:
- Magnetic resonance imaging in patients with cardiac implantable electronic devices: A prospective studyMagnetic Resonance Imaging, 2022
- Magnetic resonance imaging in all patients with implanted cardiac devices – Yes, We Can!International Journal of Cardiology, 2021
- Anatomical Model Uncertainty for RF Safety Evaluation of Metallic Implants Under MRI ExposureBioelectromagnetics, 2019
- Safety of Magnetic Resonance Imaging in Patients with Cardiac DevicesThe New England journal of medicine, 2017
- A numerical investigation on the effect of RF coil feed variability on global and local electromagnetic field exposure in human body models at 64 MHzMagnetic Resonance in Medicine, 2017
- Whole‐body and local RF absorption in human models as a function of anatomy and position within 1.5T MR body coilMagnetic Resonance in Medicine, 2013
- Safety Considerations in MRI2011
- Computational Electromagnetics: The Finite-Difference Time-Domain Method2005
- Resonant Modes of the Birdcage CoilJournal of Magnetic Resonance, 1997
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985