Design and experimental evaluation of a small-size, high-precision nuclear magnetic resonance probe for enlarged magnetic field measurement region on small-bore magnetic resonance imaging magnets
- 1 November 2026
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 97 (11)
Abstract
High-quality magnetic resonance imaging (MRI) relies on accurate magnetic field characterization; however, commercial nuclear magnetic resonance (NMR) magnetic field probes are often limited in boundary-accessible field mapping for bore-constrained magnets. In this study, a miniature free-induction-decay-based NMR magnetic field probe was developed, integrating a solenoid microcoil, a compact tuning/matching network, and a liquid sample unit, with a maximum width of ∼8 mm. Compared with a commercial NMR reference probe, the developed microprobe showed close agreement in single-point validation and yielded a mean absolute deviation of 0.27 μT in axial comparison measurements (corresponding to 0.039 ppm at 7 T), confirming its field-measurement accuracy and spatial tracking capability. Temperature-sensitivity tests over a range of 15–35 °C showed that the repeated-measurement standard deviation remained within 0.019–0.046 ppm, indicating good measurement stability. In wall-adjacent measurements inside a 172 mm-bore magnet, the maximum accessible trajectory diameter increased from 152 to 164 mm, corresponding to a 7.9% improvement in an accessible mapping range. Spherical-surface mapping of a 50 mm diameter spherical volume further yields a mean absolute deviation of 0.143 μT and a Pearson correlation coefficient greater than 0.99. These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.Keywords
Funding Information
- National Natural Science Foundation of China (52293423 and 52277031)
This publication has 29 references indexed in Scilit:
- A field camera for MR sequence monitoring and system analysisMagnetic Resonance in Medicine, 2015
- An improved RF and gradient coil system for high resolution in vivo guinea pig cochlea imaging on a 3T clinical magnetConcepts in Magnetic Resonance Part B, 2014
- Gradient system characterization by impulse response measurements with a dynamic field cameraMagnetic Resonance in Medicine, 2012
- 2H transmit-receive NMR probes for magnetic field monitoring in MRIMagnetic Resonance in Medicine, 2011
- Temperature‐induced tissue susceptibility changes lead to significant temperature errors in PRFS‐based MR thermometry during thermal interventionsMagnetic Resonance in Medicine, 2010
- NMR probes for measuring magnetic fields and field dynamics in MR systemsMagnetic Resonance in Medicine, 2008
- Spatiotemporal magnetic field monitoring for MRMagnetic Resonance in Medicine, 2008
- Solenoidal microcoil design—Part II: Optimizing winding parameters for maximum signal‐to‐noise performanceConcepts in Magnetic Resonance, 2001
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976